Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems Recall
Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems
FDA recall Class II
Ongoing
FDA lists this Class II recall for Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1112-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jan 2, 2025
- FDA report date
- Feb 12, 2025
- Quantity in commerce
- 6,994 US; 10,067 OUS
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems
Affected model & UDI checker
Search the exact identifiers published for Z-1112-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| C53987 | Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems | Check source notice | Not specified |
| Not specified | Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems | Check source notice | 15099590736002 |
FDA code information
UDI-DI 15099590736002 Lots 339062, 439811
Why was this product recalled?
Identified reagent lots are experiencing a high rate of calibration failures with failure code of "Max Iterate" on Access 2 and UniCel Dxl Immunoassay Systems. Reporting of results may be delayed due to the time required to achieve a passing calibration curve.
What actions did BECKMAN COULTER, INC. take?
An Urgent Medical Device Recall letter was sent during the week of December 30, 2024, to the affected customers via first class mail and email. If a passing calibration curve can be generated using the Access PCT reagent lots listed above, no action is needed. If repeated calibration failures with a failure code of Max Iterate are obtained using the Access PCT reagent lots listed above, discontinue using the affected reagent lot number.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1112-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology