Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1114-2026

Brand Name: Broselow Product Name: Broselow Pediatric Emergency Rainbow Tape Recall

Brand Name: Broselow Product Name: Broselow Pediatric Emergency Rainbow Tape Model/Catalog Number: 7700REA Software Version: Not Applicable Product Description: The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care. Component: No

FDA recall Class I

Ongoing

FDA lists this Class I recall for Brand Name: Broselow Product Name: Broselow Pediatric Emergency Rainbow Tape. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1114-2026
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Dec 15, 2025
FDA report date
Feb 4, 2026
Quantity in commerce
95300
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Brand Name: Broselow Product Name: Broselow Pediatric Emergency Rainbow Tape Model/Catalog Number: 7700REA Software Version: Not Applicable Product Description: The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care. Component: No

Affected model & UDI checker

Search the exact identifiers published for Z-1114-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
7700REABrand Name: Broselow Product Name: Broselow Pediatric Emergency Rainbow TapeCheck source noticeNot specified
FDA code information

Lot Code: Lot/Serial Number(s) 0004306540 0004316209 0004328895 0004329080 0004330876 0004331254 0004332994 0004338228 0004338229 0004338230 0004339717 UDI-DI Each-10889483588970 Case-30889483588974

View the original FDA recall notice

Why was this product recalled?

Product contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did SunMed Holdings, LLC take?

AirLife issued an URGENT: MEDICAL DEVICE RECALL NOTICE to its consignees on 12/15/2025 via email and US mail. The notice explained the issue, potential risk to the patient, and requested the following: "Customer immediate actions: Please take the following actions immediately: 1. Stop using the affected tapes o Immediately discontinue use of all AirLife Broselow Rainbow Tapes identified as 2025 Edition, 36-23446 Rev 3 Print Version. 2. Identify and segregate affected product o Examine your inventory and clinical areas for affected Broselow Rainbow Tapes (see Image B for identifying characteristics). o Remove all affected tapes from clinical service. o Segregate or quarantine the affected tapes to prevent further use. 3. Follow disposition instructions o Follow the instructions provided in the accompanying communication regarding discarding/destruction of the affected tapes. o Do not redistribute or place any affected tapes back into service. 4. Notify your internal users o Ensure that all clinicians and healthcare professionals within your organization who may use the Broselow Rainbow Tape are informed of this recall/field removal and understand that the affected Rev 3 tapes must not be used. 5. Notify downstream customers (if applicable) o If you have further distributed the affected tapes, please identify your customers/consignees and notify them of this recall/field removal promptly. o Your notification may be enhanced by including a copy of this letter. o If you have impacted product on hand, do not ship it; instead, hold it for discarding/destruction as instructed. 6. Complete and return the Response Form o Please complete and return the attached Response Form via e-mail to [email protected] as soon as possible. Please complete and return Attachment B if you have affected product for discarding/destruction. This allows us to document your receipt of this recall notice and the status of affected product at

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology