Brand Name: Broselow Product Name: BROSELOW" FILLED BROSELOW ORGANIZER Recall
Brand Name: Broselow Product Name: BROSELOW" FILLED BROSELOW ORGANIZER Model/Catalog Number: 7730IALS Software Version: Not Applicable Product Description: The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care. Component: no
FDA recall Class I
Ongoing
FDA lists this Class I recall for Brand Name: Broselow Product Name: BROSELOW" FILLED BROSELOW ORGANIZER. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1117-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Dec 15, 2025
- FDA report date
- Feb 4, 2026
- Quantity in commerce
- 89
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Brand Name: Broselow Product Name: BROSELOW" FILLED BROSELOW ORGANIZER Model/Catalog Number: 7730IALS Software Version: Not Applicable Product Description: The Broselow Tape is a length- and weight-based reference tool specifically designed for pediatric emergency care. Component: no
Affected model & UDI checker
Search the exact identifiers published for Z-1117-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 7730IALS | Brand Name: Broselow Product Name: BROSELOW" FILLED BROSELOW ORGANIZER | Check source notice | Not specified |
FDA code information
Lot Code: Lot/Serial Number(s) 0004317699 0004333171 0004333593 UDI-DI Each-10889483589205 Case-30889483589209
Why was this product recalled?
Product contains potentially harmful dosing errors for Vecuronium, Flumazenil, and Ketamine.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did SunMed Holdings, LLC take?
AirLife issued an URGENT: MEDICAL DEVICE RECALL NOTICE to its consignees on 12/15/2025 via email and US mail. The notice explained the issue, potential risk to the patient, and requested the following: "Customer immediate actions: Please take the following actions immediately: 1. Stop using the affected tapes o Immediately discontinue use of all AirLife Broselow Rainbow Tapes identified as 2025 Edition, 36-23446 Rev 3 Print Version. 2. Identify and segregate affected product o Examine your inventory and clinical areas for affected Broselow Rainbow Tapes (see Image B for identifying characteristics). o Remove all affected tapes from clinical service. o Segregate or quarantine the affected tapes to prevent further use. 3. Follow disposition instructions o Follow the instructions provided in the accompanying communication regarding discarding/destruction of the affected tapes. o Do not redistribute or place any affected tapes back into service. 4. Notify your internal users o Ensure that all clinicians and healthcare professionals within your organization who may use the Broselow Rainbow Tape are informed of this recall/field removal and understand that the affected Rev 3 tapes must not be used. 5. Notify downstream customers (if applicable) o If you have further distributed the affected tapes, please identify your customers/consignees and notify them of this recall/field removal promptly. o Your notification may be enhanced by including a copy of this letter. o If you have impacted product on hand, do not ship it; instead, hold it for discarding/destruction as instructed. 6. Complete and return the Response Form o Please complete and return the attached Response Form via e-mail to [email protected] as soon as possible. Please complete and return Attachment B if you have affected product for discarding/destruction. This allows us to document your receipt of this recall notice and the status of affected product at
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1117-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology