Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1124-2026

Allwell Inflation Device, for angiographic use REF: IS-30-A Recall

Allwell Inflation Device, for angiographic use REF: IS-30-A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Allwell Inflation Device, for angiographic use REF: IS-30-A. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1124-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 4, 2025
FDA report date
Jan 28, 2026
Quantity in commerce
173,645 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Allwell Inflation Device, for angiographic use REF: IS-30-A

Affected model & UDI checker

Search the exact identifiers published for Z-1124-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
IS-30-AAllwell Inflation Device, for angiographic use REF: IS-30-ACheck source noticeNot specified
Not specifiedAllwell Inflation Device, for angiographic use REF: IS-30-ACheck source notice00884450653500
FDA code information

UDI: 00884450653500 /Lot: I2559701 I2563344 I2567488 I2578111 I2593681 I2593683 I2603907 I2617788 I2622059 I2628990 I2653159 I2657407 I2853633 I2887859 I2887860 I2892046 I2917191 I2921522 I2927924 I2933535 I2989185 I2989187 I3027377 I3027378 I3037369 I3085686 I3085687 I3103552 I3118248 I3118249 I3154769 I3186052 I3186054 I3186058 I3214004 I3245239 I3245240 I3248078 I3282794

View the original FDA recall notice

Why was this product recalled?

Inflation device handle may detach from the syringe during procedure.

What actions did Merit Medical Systems, Inc. take?

On December 4, 2025 Merit Medical issued a "Urgent Medical Device Recall Notice" to affected consignees via E-Mail. Merit asked consignees to take the following actions: 1. Please read and understand the recall notification. 2. Should the handle detach during use and the balloon not completely deflate, pull the handle out of the inflation syringe, disconnect the inflation syringe from the catheter, and use an alternate syringe to generate vacuum and aspirate residual contrast 3. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them. 4. Please fill out, scan, and email the completed Customer Response Form to Customer Service at [email protected] within seven (7) calendar days

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology