BD Pyxis MedStation 4000 REF: 303. Intended to securely store and dispense medications to a qualified and authorized healthcare provider. Recall
BD Pyxis MedStation 4000 REF: 303. Intended to securely store and dispense medications to a qualified and authorized healthcare provider.
FDA recall Class II
Ongoing
FDA lists this Class II recall for BD Pyxis MedStation 4000 REF: 303. Intended to securely store and dispense medications to a qualified and authorized healthcare provider.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1125-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 23, 2024
- FDA report date
- Feb 19, 2025
- Quantity in commerce
- Not reported
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
BD Pyxis MedStation 4000 REF: 303. Intended to securely store and dispense medications to a qualified and authorized healthcare provider.
Affected model & UDI checker
Search the exact identifiers published for Z-1125-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 303 | BD Pyxis MedStation 4000 REF: 303. Intended to securely store and dispense medications to a qualified and authorized healthcare provider. | Check source notice | Not specified |
| Not specified | BD Pyxis MedStation 4000 REF: 303. Intended to securely store and dispense medications to a qualified and authorized healthcare provider. | Check source notice | 10885403512629 |
FDA code information
UDI: 10885403512629/ Serial Numbers: 16153360 16161369 16161370 16161371 16187328 16187329 16187330 15495060 15930189 16161368 16173444 16283947
Why was this product recalled?
Retroactive-Due to increase in complaints, their is a potential for drawer and/or door failures of the medication cabinet which could result in delay in accessing medications.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did CareFusion 303, Inc. take?
On February 14, 2023 and January 9. 2024 Carefusion issued a Urgent medical Device Product Advisory" Recall Notification to affected consignees. Carefusion asked consignees take the following actions: 1. Customers may continue to use these devices as intended and should file a complaint via the Customer Portal at https://www.bd.com/self-service or call BD Support at 1-800-727-6102, should you experience drawer or door failures. 2. If you experience delays in access to medication due to a drawer or door failure, ensure backup procedures (which may include but may not necessarily be limited to keys, emergency medication supplies, alternate devices in proximity, etc.) are in place and staff have all been properly trained on them. 3. Please promptly complete and return the enclosed Customer Response Form to acknowledge receipt of this notification and that you understand the information provided in this letter. 4. If you have specific questions about this letter, please contact 1-866-583-8783. 5. Any adverse health consequences experienced with the use of this product should be reported to BD and may be reported to the FDA s MedWatch Adverse Event Reporting Actions to be taken by distributor 1. Please provide a copy of this notification to any customers who may have received products listed in the Affected Products table above. 2. Please promptly complete and return the enclosed Customer Response Form to acknowledge receipt of this notification and that you understand the information provided in this letter.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1125-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology