FDA RECALL RECORD · Z-1125-2026

Merit Medical , Custom Waste Management Kit Vascular Tray REF: K10T-05261 Recall

Merit Medical , Custom Waste Management Kit Vascular Tray REF: K10T-05261

FDA recall Class II

Ongoing

FDA lists this Class II recall for Merit Medical , Custom Waste Management Kit Vascular Tray REF: K10T-05261. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1125-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 4, 2025
FDA report date
Jan 28, 2026
Quantity in commerce
864 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Merit Medical , Custom Waste Management Kit Vascular Tray REF: K10T-05261

Affected model & UDI checker

Search the exact identifiers published for Z-1125-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
K10T-05261Merit Medical , Custom Waste Management Kit Vascular Tray REF: K10T-05261Check source noticeNot specified
Not specifiedMerit Medical , Custom Waste Management Kit Vascular Tray REF: K10T-05261Check source notice00884450320297
FDA code information

UDI: 00884450320297/ Lot: T2736369 T2781930 T2898771S T3047503 T3090431 T3121937 T3146114 T3158708 T3204129 T3232415

View the original FDA recall notice

Why was this product recalled?

Inflation device handle may detach from the syringe during procedure.

What actions did Merit Medical Systems, Inc. take?

On December 4, 2025 Merit Medical issued a "Urgent Medical Device Recall Notice" to affected consignees via E-Mail. Merit asked consignees to take the following actions: 1. Please read and understand the recall notification. 2. Should the handle detach during use and the balloon not completely deflate, pull the handle out of the inflation syringe, disconnect the inflation syringe from the catheter, and use an alternate syringe to generate vacuum and aspirate residual contrast 3. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them. 4. Please fill out, scan, and email the completed Customer Response Form to Customer Service at [email protected] within seven (7) calendar days

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.