Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1129-2026

Allwell Angioplasty Pack REF: IS-30-B1/B Recall

Allwell Angioplasty Pack REF: IS-30-B1/B

FDA recall Class II

Ongoing

FDA lists this Class II recall for Allwell Angioplasty Pack REF: IS-30-B1/B. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1129-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 4, 2025
FDA report date
Jan 28, 2026
Quantity in commerce
328000 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Allwell Angioplasty Pack REF: IS-30-B1/B

Affected model & UDI checker

Search the exact identifiers published for Z-1129-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
IS-30-B1/BAllwell Angioplasty Pack REF: IS-30-B1/BCheck source noticeNot specified
Not specifiedAllwell Angioplasty Pack REF: IS-30-B1/BCheck source notice00884450653388
FDA code information

UDI:00884450653388/ Lot: I2563105 I2563106 I2565007 I2598114 I2598131 I2607891 I2612499 I2618005 I2623057 I2632527 I2632528 I2639728 I2661908 I2669772 I2669774 I2827091 I2827094 I2844360 I2844361 I2897482 I2897483 I2897484 I2897485 I2924663 I2972425 I2972426 I2972427 I2972428 I2999054 I3014683 I3030198 I3048363 I3055799 I3080949 I3089230 I3099150 I3110034 I3129711 I3129712 I3161408 I3268359 I3268361 I3268362

View the original FDA recall notice

Why was this product recalled?

Inflation device handle may detach from the syringe during procedure.

What actions did Merit Medical Systems, Inc. take?

On December 4, 2025 Merit Medical issued a "Urgent Medical Device Recall Notice" to affected consignees via E-Mail. Merit asked consignees to take the following actions: 1. Please read and understand the recall notification. 2. Should the handle detach during use and the balloon not completely deflate, pull the handle out of the inflation syringe, disconnect the inflation syringe from the catheter, and use an alternate syringe to generate vacuum and aspirate residual contrast 3. Ensure that applicable personnel within your organization are made aware of this field action. 3. If the product has been further distributed to other facilities, institutions, or manufacturers, please ensure this notice is immediately shared with them. 4. Please fill out, scan, and email the completed Customer Response Form to Customer Service at [email protected] within seven (7) calendar days

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology