Brand Name: MicroPort Product Name: HIPTURN FEM HEAD TRIAL 36MM REUSABLE Recall
Brand Name: MicroPort Product Name: HIPTURN FEM HEAD TRIAL 36MM REUSABLE Model/Catalog Number: HTHT0036 Software Version: N/A Product Description: HIPTURN FEM HEAD TRIAL 36MM REUSABLE
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: MicroPort Product Name: HIPTURN FEM HEAD TRIAL 36MM REUSABLE. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1133-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 19, 2025
- FDA report date
- Jan 28, 2026
- Quantity in commerce
- 91
- Distribution
- See key details
- FDA cause
- No Marketing Application
What is being recalled?
Brand Name: MicroPort Product Name: HIPTURN FEM HEAD TRIAL 36MM REUSABLE Model/Catalog Number: HTHT0036 Software Version: N/A Product Description: HIPTURN FEM HEAD TRIAL 36MM REUSABLE
Affected model & UDI checker
Search the exact identifiers published for Z-1133-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| HTHT0036 | Brand Name: MicroPort Product Name: HIPTURN FEM HEAD TRIAL 36MM REUSABLE | Check source notice | Not specified |
FDA code information
Lot Code: Part Number HTHT0036, all lots, no future lots planned. Primary DI Number 00192629320130. Secondary Device ID M684HTHT00361
Why was this product recalled?
Due to products not having FDA Premarket authorization to be distributed within the United States.
What actions did Microport Orthopedics, Inc. take?
On 12/19/2025, the firm email an "MicroPort Voluntary Recall - R25120001 HIPTURN Femoral Trial Head" communication to customers informing them that MicroPort was notified by the FDA that this device was not properly classified as a CLASS I device. MicroPort has decided not to pursue clearance of the device with a new FDA 510(k) submission and therefore has initiated field action to remove the device from the market. Customer are instructed to return the complete kits to MicroPort Orthopedics. For Questions - contact Bryan Callahan at 901-867-4594 or email address [email protected]
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1133-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.