Brand Name: The Guardian System Product Name: The Guardian Recall
Brand Name: The Guardian System Product Name: The Guardian Model/Catalog Number: Model AMSG3-E Product Description: The Guardian is a cardiac diagnostic implant that detects the onset of an ACS event and warns the patient to seek urgent medical care. The Guardian is an adjunct to patient recognized symptoms.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: The Guardian System Product Name: The Guardian. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1140-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Sep 1, 2023
- FDA report date
- Feb 19, 2025
- Quantity in commerce
- 548 units
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Brand Name: The Guardian System Product Name: The Guardian Model/Catalog Number: Model AMSG3-E Product Description: The Guardian is a cardiac diagnostic implant that detects the onset of an ACS event and warns the patient to seek urgent medical care. The Guardian is an adjunct to patient recognized symptoms.
Affected model & UDI checker
Search the exact identifiers published for Z-1140-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| AMSG3-E | Brand Name: The Guardian System Product Name: The Guardian | Check source notice | Not specified |
| Not specified | Brand Name: The Guardian System Product Name: The Guardian | Check source notice | 00850012625001 |
FDA code information
Model: AMSG3-E; UDI-DI: 00850012625001; Battery Lots A1946, A1947, A1948, A1949, A1950, A2003, A2004, A2031, A2032, A2033, A2034, A2035, A2036, A2038, A2039, A2040, A2041, A2042, A2043.
Why was this product recalled?
Device reaching End of Service prematurely.
What actions did Angel Medical Systems, Inc. take?
On September 18, 2023, URGENT: MEDICAL DEVICE FIELD ACTION were sent to customers. Actions to be taken by Physical/Site Avertix Medical Inc. recommends that all patients identified in the notification sheet included with this letter be checked and monitored as per the following recommendations. Patient Monitoring Recommendations: Avertix Medical Inc. recommends the following steps be taken for each such patient in this regard: 1) For the identified patients, consider reducing the interval of their 6-month regularly scheduled visits to 3-month intervals. At those visits, Avertix Medical personnel verify the battery voltage determine if it is as expected. 2) At these visits, remind patients to pay attention to the See Doctor Alerts the Guardian System Generates and to promptly notify their physician if they occur so they can be quickly evaluated. 3) Remind the patient that they can use their EXD to verify that the implanted device is working correctly by placing it up to within 2 inches of their implant location (chest) and pushing the EXD button. a. Two beeps mean the battery (and device) is functioning properly. b. One beep requires additional follow-up which includes making an appointment with their doctor to have a clinic visit so the device can be evaluated using a Physician Programmer. 4) Reinforce the fact that symptoms should never be ignored. Patients are trained at their programming visit after implant to never ignore symptoms and that training should be repeated with the patient at each scheduled visit. Acknowledging the Receipt of this Notification: Enclosed is an Acknowledgement and Receipt Form. Avertix Customer Service can assist you in completing this form and process. This information is essential to maintain field action effectiveness information required by the U.S. Food and Drug Administration. Type of Action Taken by Avertix: Avertix Medical Inc. is committed to providing the safest products possible to our physicians and patients. Once t
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1140-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology