Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1141-2025

Biological, Bio52 HealthFirst Mail In Monitoring System, 52 Tests, Item: 1003600. Biological, Bio12 HealthFirst Mail In Monitoring System, 12 Tests, Item: 1003580 Recall

Biological, Bio52 HealthFirst Mail In Monitoring System, 52 Tests, Item: 1003600. Biological, Bio12 HealthFirst Mail In Monitoring System, 12 Tests, Item: 1003580

FDA recall Class II

Ongoing

FDA lists this Class II recall for Biological, Bio52 HealthFirst Mail In Monitoring System, 52 Tests, Item: 1003600. Biological, Bio12 HealthFirst Mail In Monitoring System, 12 Tests, Item: 1003580. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1141-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 11, 2024
FDA report date
Feb 19, 2025
Quantity in commerce
4,021
Distribution
See key details
FDA cause
Error in labeling

What is being recalled?

Biological, Bio52 HealthFirst Mail In Monitoring System, 52 Tests, Item: 1003600. Biological, Bio12 HealthFirst Mail In Monitoring System, 12 Tests, Item: 1003580

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Item/UDI-DI/Lot(Expiration): 1003600/10850065007820/ 2301060(11/30/2024) 2302261(11/30/2024) 2307261(12/31/2024 2312260(1/31/2025) 2313062(1/31/2025) 2331880(6/30/2025) 2401664(8/31/2025) 2414060(2/28/2026) 2414862(3/31/2026) 2414870(3/31/2026) 2416470(3/31/2026) 2416481(4/30/2026) 2416482(4/30/2026) 1003580/10850065007813/2421181(5/31/2026)

View the original FDA recall notice

Why was this product recalled?

Spore test strip monitoring system, designed for routine monitoring of sterilization processes, may have a printing error on the return envelope that may cause spore test results to be reported to the incorrect customer's account.

What actions did HF Acquisition Co LLC take?

On 12/11/2024, correction notices were mailed to customers asking them to do the following: 1. Check the box number on your Biological Weekly Monitoring kit for the affected lot numbers. The box number is located under the bar code on the outside of the box (see image below). 3. Check the serial number on all test strip return envelopes located in the box. This number is the last set of digits in the alphanumeric string located to the left of the bar code on the envelope. 4. Verify that the box number matches the last three digits of the serial number on the return envelope. 5. If the numbers do not match, you are hereby instructed to isolate your inventory of the above-listed lot numbers and dispose of it in regular trash. Firm will work with you on replacement product. 6. Ensure this letter is promptly brought to the attention of those responsible for managing the medical device inventory. Distribute this notification to all device users within your facility and network to ensure they are informed of this field correction letter. Please share this notification with any organizations that have received the potentially affected devices. 7. Complete and return the response form via email to [email protected] Should you have any questions about this urgent field correction, please contact Customer Service at 855-476-1342 or email us at [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.