FDA RECALL RECORD · Z-1150-2025

Allura Xper FD20/20; Catalog numbers: 722038 Recall

Allura Xper FD20/20; Catalog numbers: 722038

FDA recall Class I

Ongoing

FDA lists this Class I recall for Allura Xper FD20/20; Catalog numbers: 722038. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1150-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Jan 13, 2025
FDA report date
Feb 26, 2025
Quantity in commerce
114 units (33 US, 81 OUS)
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Allura Xper FD20/20; Catalog numbers: 722038

Affected model & UDI checker

Search the exact identifiers published for Z-1150-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
722038Allura Xper FD20/20; Catalog numbers: 722038Check source noticeNot specified
Not specifiedAllura Xper FD20/20; Catalog numbers: 722038Check source notice00884838054226
FDA code information

Catalog number: 722038; UDI-DI: 00884838054226; Serial Numbers: 39, 63, 42, 55, 67, 99, 60, 71, 141, 115, 7, 78, 108, 106, 167, 118, 130, 49, 119, 126, 173, 37, 38, 48, 62, 35, 137, 120, 41, 157, 96, 70, 46, 102, 153, 155, 171, 172, 163, 129, 161, 122, 156, 121, 168, 113, 169, 160, 152, 162, 164, 148, 159, 144, 111, 138, 158, 124, 140, 154, 147, 14, 166, 127, 94, 95, 90, 91, 93, 143, 92, 89, 150, 88, 34, 136, 66, 30, 58, 29, 54, 170, 28, 47, 36, 79, 98, 31, 146, 3, 123, 32, 5, 9, 27, 105, 69, 40, 75, 11, 104, 112, 50, 116, 73, 133, 110, 72, 117, 134, 135, 68, 64, 45, 109.

View the original FDA recall notice

Why was this product recalled?

Possibility of the patient falling from the table related to the mattress used on the patient table.

What actions did PHILIPS MEDICAL SYSTEMS NEDERLAND B.V. take?

On January 14, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. 4. Actions that should be taken by the customer / user in order to prevent risks for patients or users " Circulate this Urgent Medical Device Correction letter to all users of the system so that they are aware of the issue. " Follow the additional instructions on the use of the mattress included in Appendix B. These instructions include information for the transfer of the patient and the correct use and positioning of the mattress. The information of this Appendix is being included in an Addendum to the Instructions for Use. Additionally, Philips is creating a Quick Reference Card. The Addendum and the Quick Reference Card will be available as from end of March 2025. Please refer to Appendix C on how to download the Addendum to the Instructions for Use and the Quick Reference Card. " In case that the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through the email account: [email protected] " Please complete and return the Response form attached to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Urgent Medical Device Correction letter and understanding of the issue and required actions to be taken. " Should you experience the issue reported in this letter, please report the event to Philips through your local Philips representative (1- 800-722-9377). 5. Actions planned by Philips Image Guided Therapy Systems to correct the problem Philips is informing customers of this issue through this Urgent Medical Device Correction. Additionally, Philips is releasing an Addendum to the Instructions for Use and a Quick Reference Card that will be available for customers at the end of March 2025. This Urgent Medical Device Correction letter has been reported to the appropriate Regulatory Agencies.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history