Brand Name: Elekta Medical Linear Accelerator Product Name: Elekta Synergy, Elekta Harmony, Elekta Infinity, Versa HD Software Version: N/A Product Description: Digital Linear Accelerator Component: No Recall
Brand Name: Elekta Medical Linear Accelerator Product Name: Elekta Synergy, Elekta Harmony, Elekta Infinity, Versa HD Software Version: N/A Product Description: Digital Linear Accelerator Component: No
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Elekta Medical Linear Accelerator Product Name: Elekta Synergy, Elekta Harmony, Elekta Infinity, Versa HD Software Version: N/A Product Description: Digital Linear Accelerator Component: No. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1164-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 14, 2025
- FDA report date
- Feb 4, 2026
- Quantity in commerce
- 5391
- Distribution
- See key details
- FDA cause
- Process control
What is being recalled?
Brand Name: Elekta Medical Linear Accelerator Product Name: Elekta Synergy, Elekta Harmony, Elekta Infinity, Versa HD Software Version: N/A Product Description: Digital Linear Accelerator Component: No
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
Lot Code: 05060191071550 (NFB) Elekta Synergy 05060191071505 (FB) 05060191071543 (NFB) Elekta Infinity 05060191071512 (FB) 05060191071574 (NFB) Versa HD 05060191071529 (FB) 05060191071581 Harmony 05060191071604 Harmony Pro 07340201503461 Elekta Evo
Why was this product recalled?
Due to a manufacturing issue and electrical grounding of systems.
What actions did Elekta, Inc. take?
On 12/14/2025, the firm emailed an "URGENT IMPORTANT FIELD SAFETY NOTIFICATION" Letter to customer informing them that Elekta has identified a safety issue relating to the modulator discharge rod, where the supplier has been fitting the crimp in the crimp tool back to front. The issue was discovered during production. Note - The discharge rod is used by Field Service Engineers (FSE) when completing service activities on the machine around the HT and RF (High Tension (high voltage) and Radio Frequency) assembly. Instructions to customer: No actions are required to be taken by the Clinical user. Instructions for Field Service Engineers: Possible impact on the Field Service Engineer:- The FSE uses the HT discharge rod to fully discharge the voltage capacitor. Once the FSE has completed the procedure using the discharge rod, the FSE will start to work on the system around the HT and RF assembly. If the discharge rod is not fully connected to the ground or not connected at all, the system will still be fully charged and the FSE could receive an electric shock. Elekta will release an Important Field Safety Modification (IFSM) to identify potentially faulty discharge rod crimping and provide further instructions. The IFSM will include procedures for verifying the reliability of the discharge rod crimping and replacing affected components if necessary. local Elekta service representative will contact you to arrange for this activity to be completed.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1164-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026