Genesis Screw-In Anchor Recall

Genesis Screw-In Anchor

FDA recall Class II

Ongoing

FDA lists this Class II recall for Genesis Screw-In Anchor. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1168-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 17, 2025
FDA report date
Feb 4, 2026
Quantity in commerce
600 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Genesis Screw-In Anchor

Affected model & UDI checker

Search the exact identifiers published for Z-1168-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
10B45003Genesis Screw-In AnchorCheck source noticeNot specified
FDA code information

Model No. 10B45003, 10B55003, 10B65003; All UDI Codes; All Lots.

View the original FDA recall notice

Why was this product recalled?

Reports of device failure, including anchor breakage, anchor bending, anchor pull out, suture breakage, and secondary packaging breakage.

What actions did Aju Pharm Co., Ltd. take?

An URGENT MEDICAL DEVICE RECALL notice dated 12/17/25 was mailed to consignees. The notification instructs consignees to discontinue use of affected devices, segregate affected devices, and return the product to Aju Pharm Medical or the authorized distributor. If product was further distributed the recall notification is to be forwarded for their awareness. Aju Pharm states that no additional action is required for patients who have already undergone surgery using affected devices. Consignees with any questions should contact +82-02-2630-0641 by phone or [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology