Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1171-2025

Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A136 used in conjunction with Warming Blankets and Mattresses. Recall

Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A136 used in conjunction with Warming Blankets and Mattresses.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A136 used in conjunction with Warming Blankets and Mattresses.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1171-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 15, 2024
FDA report date
Feb 26, 2025
Quantity in commerce
20 units
Distribution
See key details
FDA cause
Device Design

What is being recalled?

Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A136 used in conjunction with Warming Blankets and Mattresses.

Affected model & UDI checker

Search the exact identifiers published for Z-1171-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
A136Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A136 used in conjunction with Warming Blankets and Mattresses.Check source noticeNot specified
Not specifiedAugustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A136 used in conjunction with Warming Blankets and Mattresses.Check source notice00855913001794
FDA code information

UDI-DI 00855913001794, Lot Number 2122

View the original FDA recall notice

Why was this product recalled?

Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .

What actions did Augustine Temperature Management, LLC take?

Augustine Temperature Management contacted their consignees by visit beginning on 02/15/2024 to replace the affected cables. The firm collected email confirmation that the original cables had either been returned or destroyed. If you have any questions, contact the Product Manager at 952-465-3526 or email:

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.