Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A138 used in conjunction with Warming Blankets and Mattresses. Recall
Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A138 used in conjunction with Warming Blankets and Mattresses.
FDA recall Class II
Ongoing
FDA lists this Class II recall for Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A138 used in conjunction with Warming Blankets and Mattresses.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1172-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 15, 2024
- FDA report date
- Feb 26, 2025
- Quantity in commerce
- 7 unit
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A138 used in conjunction with Warming Blankets and Mattresses.
Affected model & UDI checker
Search the exact identifiers published for Z-1172-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| A138 | Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A138 used in conjunction with Warming Blankets and Mattresses. | Check source notice | Not specified |
| Not specified | Augustine Surgical Inc. HotDog Return Electrode accessory connector cables, Model A138 used in conjunction with Warming Blankets and Mattresses. | Check source notice | 00850011479674 |
FDA code information
UDI-DI 00850011479674, Lot Number 2339
Why was this product recalled?
Users were experiencing the following error message (M-0B-125) stated Check Neutral Electrode Contact .
What actions did Augustine Temperature Management, LLC take?
Augustine Temperature Management contacted their consignees by visit beginning on 02/15/2024 to replace the affected cables. The firm collected email confirmation that the original cables had either been returned or destroyed. If you have any questions, contact the Product Manager at 952-465-3526 or email:
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1172-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.