DeRoyal AMG Suction Connector Tubing REF: 71-2083, 71-C506, 71-C510, 71-C512, 71-C520, Tubing is intended for fluid movement Recall
DeRoyal AMG Suction Connector Tubing REF: 71-2083, 71-C506, 71-C510, 71-C512, 71-C520, Tubing is intended for fluid movement
FDA recall Class II
Ongoing
FDA lists this Class II recall for DeRoyal AMG Suction Connector Tubing REF: 71-2083, 71-C506, 71-C510, 71-C512, 71-C520, Tubing is intended for fluid movement. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1174-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jan 15, 2025
- FDA report date
- Feb 26, 2025
- Quantity in commerce
- 224,770 tubes/connectors
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
DeRoyal AMG Suction Connector Tubing REF: 71-2083, 71-C506, 71-C510, 71-C512, 71-C520, Tubing is intended for fluid movement
Affected model & UDI checker
Search the exact identifiers published for Z-1174-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 71-2083 | DeRoyal AMG Suction Connector Tubing REF: 71-2083, 71-C506, 71-C510, 71-C512, 71-C520, Tubing is intended for fluid movement | Check source notice | Not specified |
| Not specified | DeRoyal AMG Suction Connector Tubing REF: 71-2083, 71-C506, 71-C510, 71-C512, 71-C520, Tubing is intended for fluid movement | Check source notice | 00749756578981 |
| Not specified | DeRoyal AMG Suction Connector Tubing REF: 71-2083, 71-C506, 71-C510, 71-C512, 71-C520, Tubing is intended for fluid movement | Check source notice | 00749756579032 |
| Not specified | DeRoyal AMG Suction Connector Tubing REF: 71-2083, 71-C506, 71-C510, 71-C512, 71-C520, Tubing is intended for fluid movement | Check source notice | 00749756579063 |
| Not specified | DeRoyal AMG Suction Connector Tubing REF: 71-2083, 71-C506, 71-C510, 71-C512, 71-C520, Tubing is intended for fluid movement | Check source notice | 00749756579070 |
FDA code information
71-2083 Connector Tubing 71-2084 71-C501 Suction Tubing, Connector Item Number: 548501 UDI-DI code: 00749756578981 Lot Numbers: 71-C506 Tubing Suction Connector Item Number: 548527 UDI-DI code: 00749756579032 Lot Numbers: CNWKD04-01 exp 4/4/2029 CWNKD04-02 exp 4/9/2029 CNWKD04-03 exp 4/14/2029 CNWKD04-04 exp 4/19/2029 CNWKD04-05 exp 4/24/2029 CNWKD04-06 exp 4/29/2029 CNWKD04-07 exp 5/4/2029 CNWKD04-08 exp 5/9/2029 CNWKD04-09 exp 5/14/2029 CNWKD04-10 exp 5/19/2029 CNWKD09-02 exp 9/1/2029 CNWKD09-08 exp 9/13/2029 71-C510 Tubing Suction Connector Item Number: 548535 UDI-DI code: '00749756579049 Lot Numbers: CNWKD04-01 exp 4/4/2029 CNWKD04-02 exp 4/9/2029 CNWKD04-03 exp 4/14/2029 CNWKD04-04 exp 4/19/2024 CNWKD04-05 exp 4/24/2029 CNWKD04-06 exp 4/29/2029 CNWKD04-07 exp 5/4/2029 CNWKD04-08 exp 5/9/2029 CNWKD04-09 exp 5/14/2029 CNWKD04-10 exp 5/19/2029 71-C512 Tubing Suction Connector Item Number: 548543 UDI-DI code: 00749756579063 Lot Numbers: CNWKD04-01 exp 4/4/2029 CNWKD04-02 exp 4/9/2029 CNWKD04-03 exp 4/14/2029 CNWKD04-04 exp 4/19/2029 CNWKD04-05 exp 4/24/2029 CNWKD04-06 exp 4/29/2029 CNWKD04-07 exp 5/4/2029 CNWKD04-08 exp 5/9/2029 CNWKD04-09 exp 5/14/2029 CNWKD04-10 exp 5/19/2029 CNWKD09-05 exp 9/7/2029 CNWKD09-07 exp 9/11/2029 CNWKD09-08 exp 9/13/2029 CNWKD10-05 exp 10/9/2029 71-C520 Tubing Suction Connector Item Number: 548551 UDI-DI code: 00749756579070 Lot Numbers: CNWKD04-01 exp 4/4/2029 CNWKD04-04 exp 4/19/2029 CNWKD09-02 exp 9/1/2029 CNWKD10-06 exp 10/12/2029
Why was this product recalled?
The tubing may not ensure that the yankauer stays intact, and secure to the tubing / yankauer connection for a tight seal. This could result in inadequate suction and cause delay in patient care.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did DEROYAL INDUSTRIES, INC. take?
On 01/15/2025 the firm sent via email an "URGENT! DEROYAL RECALL NOTICE" letter for the Suction Connector Tubing. The letter was informing customers that the suction connector tubing could be defective and that the tubing may not ensure that the yankauer stays intact, and secure to the tubing/yankauer connection for a tight seal. Customers are instructed to: 1.Search their inventory for the affected products and place them in quarantine to prevent further use. 2. Destroy all affected products and complete the NOTICE OF RETURN FORM and indicate all affected products found in their inventory. Return completed form to DeRoyal via fax to 865-362-3716 or [email protected]. 3. If customers have questions regarding credit for the recall, contact Jennifer Marsh at [email protected] or by phone 865-362-4203 or email at [email protected]. If replacement product is preferred, contact DeRoyal Sales Representative or contact Angie Schubert, Assistant Product Director at [email protected]
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1174-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology