AnchorFast Guard Select, Oral Endotracheal Tube Fastener, REF 9700 Recall
AnchorFast Guard Select, Oral Endotracheal Tube Fastener, REF 9700,
FDA recall Class II
Ongoing
FDA lists this Class II recall for AnchorFast Guard Select, Oral Endotracheal Tube Fastener, REF 9700. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1178-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 23, 2024
- FDA report date
- Feb 26, 2025
- Quantity in commerce
- 701 box / 8,412 eaches
- Distribution
- See key details
- FDA cause
- Process change control
What is being recalled?
AnchorFast Guard Select, Oral Endotracheal Tube Fastener, REF 9700,
Affected model & UDI checker
Search the exact identifiers published for Z-1178-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 9700 | AnchorFast Guard Select, Oral Endotracheal Tube Fastener, REF 9700 | Check source notice | Not specified |
| Not specified | AnchorFast Guard Select, Oral Endotracheal Tube Fastener, REF 9700 | Check source notice | 00610075095810 |
FDA code information
UDI/DI 00610075095810, Lot Numbers: 4J162, 4J172, 4J182, 4L172, 4L182
Why was this product recalled?
Hollister received reports of decreased skin barrier wear time which could can lead to tube migration
What actions did Hollister Incorporated take?
Hollister issued an URGENT - MEDICAL DEVICE RECALL notice to its consignees on 12/23/2024 via FedEx and email. The notice explained the reason for the recall and potential risk and instructed customers to quarantine affected inventory, discard the product per their local procedures for product destruction, and Please complete and return the response form attached as soon as possible, even if you do not have affected product in your possession and send to [email protected]. For product credit requests: If you are a direct customer of Hollister, contact Hollister Incorporated Customer Service for product credit at: 1-800-323-4060 (prompt #1). If you purchase product from a distributor, please contact your distributor for product credit. Hollister has posted this recall on their website with the link to the page.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1178-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.