Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1184-2025

LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo LNCS-II rainbow DCI 8 SpCO, Adult Reusable Sensor REF 11996-000519 Masimo LNCS-II rainbow DCI 8 SpCO, Pediatric Reusable Sensor REF 11996-000520 The LIFEPAK 15 Monitor/Defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. RD rainbow" 8 SpCO sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), and methemoglobin saturation (SpMet) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments. Recall

LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo¿ RD rainbow Adt 8¿ SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo¿ LNCS-II rainbow¿ DCI 8¿ SpCO, Adult Reusable Sensor REF 11996-000519 Masimo¿ LNCS-II rainbow¿ DCI 8¿ SpCO, Pediatric Reusable Sensor REF 11996-000520 The LIFEPAK¿ 15 Monitor/Defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. RD rainbow" 8¿ SpCO sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO¿), and methemoglobin saturation (SpMet¿) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.

FDA recall Class II

Ongoing

FDA lists this Class II recall for LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo LNCS-II rainbow DCI 8 SpCO, Adult Reusable Sensor REF 11996-000519 Masimo LNCS-II rainbow DCI 8 SpCO, Pediatric Reusable Sensor REF 11996-000520 The LIFEPAK 15 Monitor/Defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. RD rainbow" 8 SpCO sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), and methemoglobin saturation (SpMet) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1184-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 21, 2025
FDA report date
Feb 26, 2025
Quantity in commerce
622
Distribution
See key details
FDA cause
Device Design

What is being recalled?

LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo LNCS-II rainbow DCI 8 SpCO, Adult Reusable Sensor REF 11996-000519 Masimo LNCS-II rainbow DCI 8 SpCO, Pediatric Reusable Sensor REF 11996-000520 The LIFEPAK 15 Monitor/Defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. RD rainbow" 8 SpCO sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), and methemoglobin saturation (SpMet) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.

Affected model & UDI checker

Search the exact identifiers published for Z-1184-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
11996-000515LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo LNCS-II rainbow DCI 8 SpCO, Adult Reusable Sensor REF 11996-000519 Masimo LNCS-II rainbow DCI 8 SpCO, Pediatric Reusable Sensor REF 11996-000520 The LIFEPAK 15 Monitor/Defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. RD rainbow" 8 SpCO sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), and methemoglobin saturation (SpMet) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.Check source noticeNot specified
11996-000519LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo LNCS-II rainbow DCI 8 SpCO, Adult Reusable Sensor REF 11996-000519 Masimo LNCS-II rainbow DCI 8 SpCO, Pediatric Reusable Sensor REF 11996-000520 The LIFEPAK 15 Monitor/Defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. RD rainbow" 8 SpCO sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), and methemoglobin saturation (SpMet) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.Check source noticeNot specified
11996-000520LIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo LNCS-II rainbow DCI 8 SpCO, Adult Reusable Sensor REF 11996-000519 Masimo LNCS-II rainbow DCI 8 SpCO, Pediatric Reusable Sensor REF 11996-000520 The LIFEPAK 15 Monitor/Defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. RD rainbow" 8 SpCO sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), and methemoglobin saturation (SpMet) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.Check source noticeNot specified
Not specifiedLIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo LNCS-II rainbow DCI 8 SpCO, Adult Reusable Sensor REF 11996-000519 Masimo LNCS-II rainbow DCI 8 SpCO, Pediatric Reusable Sensor REF 11996-000520 The LIFEPAK 15 Monitor/Defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. RD rainbow" 8 SpCO sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), and methemoglobin saturation (SpMet) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.Check source notice00843997010863
Not specifiedLIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo LNCS-II rainbow DCI 8 SpCO, Adult Reusable Sensor REF 11996-000519 Masimo LNCS-II rainbow DCI 8 SpCO, Pediatric Reusable Sensor REF 11996-000520 The LIFEPAK 15 Monitor/Defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. RD rainbow" 8 SpCO sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), and methemoglobin saturation (SpMet) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.Check source notice00843997015608
Not specifiedLIFEPAK 15 V4/Masimo Rainbow Sensors: Masimo RD rainbow Adt 8 SpCO, SpO2, and SpMet, Adult Adhesive Sensors REF 11996-000515 Masimo LNCS-II rainbow DCI 8 SpCO, Adult Reusable Sensor REF 11996-000519 Masimo LNCS-II rainbow DCI 8 SpCO, Pediatric Reusable Sensor REF 11996-000520 The LIFEPAK 15 Monitor/Defibrillator (LP15) is a complete acute cardiac care response system designed for basic life support (BLS) and advanced life support (ALS) patient management protocols. RD rainbow" 8 SpCO sensors are indicated for the continuous noninvasive monitoring of functional oxygen saturation of arterial hemoglobin (SpO2), pulse rate, carboxyhemoglobin saturation (SpCO), and methemoglobin saturation (SpMet) for use with adult, pediatric, infant, and neonatal patients during both no motion and motion conditions, and for patients who are well or poorly perfused in hospitals, hospital-type facilities, mobile, and home environments.Check source notice00843997015615
FDA code information

Catalog Number: 11996-000515 UDI-DI/GTIN code: 00843997010863 Lot Number: 23G79 Catalog Number: 11996-000519 UDI-DI/GTIN code: 00843997015608 Lot Numbers: 23HER 23HPT 23JBN 23JTA Catalog Number: 11996-000520 UDI-DI/GTIN code: 00843997015615 Lot Numbers: 23HNV 23JBP

View the original FDA recall notice

Why was this product recalled?

Due to an error message that prevents users from utilizing carboxyhemoglobin saturation and methemoglobin saturation sensors on their monitor/defibrillator system.

What actions did Physio-Control, Inc. take?

On 1/21/2025, the firm send an "URGENT: MEDICAL DEVICE RECALL" Letter via UPS 2-Day mail to customers to inform them that the LIFEPAK 15 V4 displays an error where "SpO2: Sensor does not support SpCO or SpMET". The error results in preventing users from utilizing LIFEPAK 15 V4 SpCO and SpMet functionalities when connected to specific lots of Masimo Rainbow Sensors. Customers are instructed to: 1. Immediately check your internal inventory to locate the product listed on the attached business reply form and remove them from their point of use. 2. Return the enclosed business reply form by email to [email protected] to confirm receipt of this notification. 3. Upon receipt of the completed business reply form, Stryker will contact you to arrange for the return and replacement of your product(s). 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. Please keep this communication close to the affected items until replacement. 5. Inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients appropriately. For questions or concerns - contact Customer Service +1 800 787 9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time), Monday Friday or by email at [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.