Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1195-2026

Medline medical procedure convenience kits labeled as: 1) INPATIENT CTRL LN DRG CHG, Kit SKU DT22880; 2) ADULT CVC DRSG CHG W/ADHESIVE, Kit SKU DYNDC3283A. Recall

Medline medical procedure convenience kits labeled as: 1) INPATIENT CTRL LN DRG CHG, Kit SKU DT22880; 2) ADULT CVC DRSG CHG W/ADHESIVE, Kit SKU DYNDC3283A.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Medline medical procedure convenience kits labeled as: 1) INPATIENT CTRL LN DRG CHG, Kit SKU DT22880; 2) ADULT CVC DRSG CHG W/ADHESIVE, Kit SKU DYNDC3283A.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1195-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 12, 2025
FDA report date
Feb 4, 2026
Quantity in commerce
1,928 total
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Medline medical procedure convenience kits labeled as: 1) INPATIENT CTRL LN DRG CHG, Kit SKU DT22880; 2) ADULT CVC DRSG CHG W/ADHESIVE, Kit SKU DYNDC3283A.

Affected model & UDI checker

Search the exact identifiers published for Z-1195-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMedline medical procedure convenience kits labeled as: 1) INPATIENT CTRL LN DRG CHG, Kit SKU DT22880; 2) ADULT CVC DRSG CHG W/ADHESIVE, Kit SKU DYNDC3283A.Check source notice10653160995470
Not specifiedMedline medical procedure convenience kits labeled as: 1) INPATIENT CTRL LN DRG CHG, Kit SKU DT22880; 2) ADULT CVC DRSG CHG W/ADHESIVE, Kit SKU DYNDC3283A.Check source notice40653160995471
Not specifiedMedline medical procedure convenience kits labeled as: 1) INPATIENT CTRL LN DRG CHG, Kit SKU DT22880; 2) ADULT CVC DRSG CHG W/ADHESIVE, Kit SKU DYNDC3283A.Check source notice10198459313240
Not specifiedMedline medical procedure convenience kits labeled as: 1) INPATIENT CTRL LN DRG CHG, Kit SKU DT22880; 2) ADULT CVC DRSG CHG W/ADHESIVE, Kit SKU DYNDC3283A.Check source notice40198459313241
FDA code information

Medline Kit SKU DT22880: UDI/DI 10653160995470 (EA) 40653160995471 (CS), Lot Number 25KMJ441; Medline Kit SKU DYNDC3283A: UDI/DI 10198459313240 (EA) 40198459313241 (CS), Lot Number 25KMH515.

View the original FDA recall notice

Why was this product recalled?

Medline kits were manufactured and distributed containing MASTISOL liquid adhesive which was recalled due to complaints that the butyrate tubing cracked during actuation of the vials of MASTISOL Liquid Adhesive.

What actions did Medline Industries, LP take?

Medline Industries issued a URGENT MEDICAL DEVICE RECALL notice to its consignees on 12/12/2025 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1. Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2. Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com Recall Reference #: R-25-251-FGX2 Recall Code: 3. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. 4. If you are a distributor or have resold or transferred this product to another company or individual, you are required by FDA regulations to notify them of this recall communication. Please include your customers quantities on your response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product." For questions, contact the Recall Department at 866-359-1704 or [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology
Z-1195-2026: Medline medical procedure convenience kits labeled as: 1) INPATIENT CTRL LN DRG CHG, Kit SKU DT recall | MedtechLedger