Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia Recall
Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia
FDA recall Class II
Ongoing
FDA lists this Class II recall for Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1197-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Dec 16, 2024
- FDA report date
- Feb 26, 2025
- Quantity in commerce
- 4 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia
Affected model & UDI checker
Search the exact identifiers published for Z-1197-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| US501 | Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia | Check source notice | Not specified |
| Not specified | Hot Dog PATIENT WARMING, Underbody Warming Mattress + Return Electrode, 82 cm (32"), REF US501, to prevent or treat hypothermia | Check source notice | 00855913001817 |
FDA code information
UDI-DI UDI-DI 00855913001817, Serial Numbers: 1304292, 1304293, 1304296, 1304297
Why was this product recalled?
There is over-molding that stabilizes the connection between the cable and cable-entry on the mattress. It turns out that after about a year of use, the connector cable on the mattress can experience the same type of deflection that caused the intermittent connection with the cables connecting to the ESU.
What actions did Augustine Temperature Management, LLC take?
Augustine Temperature Management began notifying its consignees on 12/16/2024 via email and visit to exchange the affected product.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1197-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.