Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1202-2026

Brand Name: Zimmer Tourniquet Systems Product Name: A.T.S. 3200TS Tourniquet Systems Recall

Brand Name: Zimmer Tourniquet Systems Product Name: A.T.S.¿ 3200TS Tourniquet Systems Model/Catalog Number: 60320010100 Software Version: Software version v2.06 or prior are within scope of the recall for A.T.S.¿ 3200TS Product Description: A.T.S 3200TS Tourniquet Systems Component: N/A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Zimmer Tourniquet Systems Product Name: A.T.S. 3200TS Tourniquet Systems. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1202-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 24, 2025
FDA report date
Feb 4, 2026
Quantity in commerce
51
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Brand Name: Zimmer Tourniquet Systems Product Name: A.T.S. 3200TS Tourniquet Systems Model/Catalog Number: 60320010100 Software Version: Software version v2.06 or prior are within scope of the recall for A.T.S. 3200TS Product Description: A.T.S 3200TS Tourniquet Systems Component: N/A

Affected model & UDI checker

Search the exact identifiers published for Z-1202-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
60320010100Brand Name: Zimmer Tourniquet Systems Product Name: A.T.S. 3200TS Tourniquet SystemsCheck source noticeNot specified
FDA code information

Lot Code: Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739001" ; Lot No. 67505739 ; Serial No. 32067505739001 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739003" ; Lot No. 67505739 ; Serial No. 32067505739003 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739004" ; Lot No. 67505739 ; Serial No. 32067505739004 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739005" ; Lot No. 67505739 ; Serial No. 32067505739005 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739006" ; Lot No. 67505739 ; Serial No. 32067505739006 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739007" ; Lot No. 67505739 ; Serial No. 32067505739007 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739008" ; Lot No. 67505739 ; Serial No. 32067505739008 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739010" ; Lot No. 67505739 ; Serial No. 32067505739010 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739013" ; Lot No. 67505739 ; Serial No. 32067505739013 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739014" ; Lot No. 67505739 ; Serial No. 32067505739014 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739015" ; Lot No. 67505739 ; Serial No. 32067505739015 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739017" ; Lot No. 67505739 ; Serial No. 32067505739017 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739019" ; Lot No. 67505739 ; Serial No. 32067505739019 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67505739(21)32067505739020" ; Lot No. 67505739 ; Serial No. 32067505739020 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816001" ; Lot No. 67508816 ; Serial No. 32067508816001 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816002" ; Lot No. 67508816 ; Serial No. 32067508816002 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816003" ; Lot No. 67508816 ; Serial No. 32067508816003 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816004" ; Lot No. 67508816 ; Serial No. 32067508816004 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816005" ; Lot No. 67508816 ; Serial No. 32067508816005 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816006" ; Lot No. 67508816 ; Serial No. 32067508816006 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816007" ; Lot No. 67508816 ; Serial No. 32067508816007 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816008" ; Lot No. 67508816 ; Serial No. 32067508816008 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816009" ; Lot No. 67508816 ; Serial No. 32067508816009 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816010" ; Lot No. 67508816 ; Serial No. 32067508816010 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816012" ; Lot No. 67508816 ; Serial No. 32067508816012 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816013" ; Lot No. 67508816 ; Serial No. 32067508816013 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816014" ; Lot No. 67508816 ; Serial No. 32067508816014 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816016" ; Lot No. 67508816 ; Serial No. 32067508816016 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816017" ; Lot No. 67508816 ; Serial No. 32067508816017 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816019" ; Lot No. 67508816 ; Serial No. 32067508816019 Model No. 60320010100 ; UDI "(01)00889024626447(11)250725(10)67508816(21)32067508816020" ; Lot No. 67508816 ; Serial No. 32067508816020 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820001" ; Lot No. 67508820 ; Serial No. 32067508820001 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820002" ; Lot No. 67508820 ; Serial No. 32067508820002 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820003" ; Lot No. 67508820 ; Serial No. 32067508820003 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820004" ; Lot No. 67508820 ; Serial No. 32067508820004 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820005" ; Lot No. 67508820 ; Serial No. 32067508820005 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820006" ; Lot No. 67508820 ; Serial No. 32067508820006 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820007" ; Lot No. 67508820 ; Serial No. 32067508820007 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820008" ; Lot No. 67508820 ; Serial No. 32067508820008 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820009" ; Lot No. 67508820 ; Serial No. 32067508820009 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820010" ; Lot No. 67508820 ; Serial No. 32067508820010 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820011" ; Lot No. 67508820 ; Serial No. 32067508820011 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820012" ; Lot No. 67508820 ; Serial No. 32067508820012 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820013" ; Lot No. 67508820 ; Serial No. 32067508820013 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820014" ; Lot No. 67508820 ; Serial No. 32067508820014 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820015" ; Lot No. 67508820 ; Serial No. 32067508820015 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820016" ; Lot No. 67508820 ; Serial No. 32067508820016 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820017" ; Lot No. 67508820 ; Serial No. 32067508820017 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820018" ; Lot No. 67508820 ; Serial No. 32067508820018 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820019" ; Lot No. 67508820 ; Serial No. 32067508820019 Model No. 60320010100 ; UDI "(01)00889024626447(11)250730(10)67508820(21)32067508820020" ; Lot No. 67508820 ; Serial No. 32067508820020

View the original FDA recall notice

Why was this product recalled?

Device user interface (screen) may freeze or become unresponsive to touch when operating in certain non-English language settings. Issue may cause delay or extension while alternate device is located.

What actions did Zimmer Surgical Inc take?

On December 30, 2025, the firm began notifying consignees through Urgent Medical Device Recall letters. The firm is conducting this recall in two phases. In Phase 1, which began on 12/30/2025, consignees are informed that the screen may freeze or stop responding to touch when used in certain international languages. Users may continue to use the device by utilizing the English language mode. Use of the device in an affected language is not advised, and may cause the issue to occur. ATS5000 Affected Languages: Czech, Danish, Finnish, French, Italian, Lithuanian, Norwegian, Romanian, Slovak, Slovenian, and Turkish Unaffected Languages: English, Chinese (simplified), Korean, Japanese, Greek, Bulgarian, and Russian All devices manufactured between April 2024 - September 2025 with software version v2.04 are within scope of the recall. Affected devices will require an upgrade to the new software version v2.05 when it is available in early 2026. The firm will issue a removal notice (Phase 2) once the new software version is available. Customers are instructed to not return affected devices before the initiation of Phase 2.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology