FDA RECALL RECORD · Z-1208-2025

IntelePACS (Image Fusion Module) - InteleViewer Recall

IntelePACS (Image Fusion Module) - InteleViewer

FDA recall Class II

Ongoing

FDA lists this Class II recall for IntelePACS (Image Fusion Module) - InteleViewer. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1208-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 5, 2025
FDA report date
Mar 5, 2025
Quantity in commerce
88
Distribution
See key details
FDA cause
Software design

What is being recalled?

IntelePACS (Image Fusion Module) - InteleViewer

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

UDI-DI: B228INTELEPACS0, Device Versions: 5-6-1-P23 and later

View the original FDA recall notice

Why was this product recalled?

A software application that receives digital images and data from various sources has a bug in specific software versions that will calculate the Standard Uptake Value (SUV) incorrectly for PET/CT studies acquired on the days of Daylight Savings Time (DST); only on dates in which DST transition occurs; all other dates are unaffected. An incorrect SUV may lead to an incorrect diagnosis.

What actions did INTELERAD MEDICAL SYSTEMS INCORPORATED take?

On 2/5/2025, correction notices were emailed to customers who were asked to do the following: The following work around can be completed for any impacted SUV calculation. The correct SUV values may be obtained in the SUV calculator, as follows: 1. Adjust the administration time and series time in IntelePACS consistent with the actual time of injection and the time of series acquisition. 2. Verify that the calculated elapsed time is correct. 3. Recalculate the SUV in the calculator, this is the correct SUV value. - Customers are asked to review scans and notify the recalling firm if scans were performed on Daylight Savings Time dates since 2022. - A software upgrade to eliminate this bug is expected to be released by the end of March 2025. Once available, you will be notified via email. You will then be able to schedule your product upgrade. - Complete and return the acknowledgement form via email to [email protected] For more information or to ask any questions about this notification, please contact the firm's regulatory affairs department at [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.