FDA RECALL RECORD · Z-1210-2026

Ingenia Ambition X. Product Code (REF): (1) 781356, (2) 782109, (3)782138. MR systems with SW version R11.1 and R12.1. Recall

Ingenia Ambition X. Product Code (REF): (1) 781356, (2) 782109, (3)782138. MR systems with SW version R11.1 and R12.1.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Ingenia Ambition X. Product Code (REF): (1) 781356, (2) 782109, (3)782138. MR systems with SW version R11.1 and R12.1.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1210-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 3, 2025
FDA report date
Feb 4, 2026
Quantity in commerce
21 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

Ingenia Ambition X. Product Code (REF): (1) 781356, (2) 782109, (3)782138. MR systems with SW version R11.1 and R12.1.

Affected model & UDI checker

Search the exact identifiers published for Z-1210-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedIngenia Ambition X. Product Code (REF): (1) 781356, (2) 782109, (3)782138. MR systems with SW version R11.1 and R12.1.Check source notice00884838090040
Not specifiedIngenia Ambition X. Product Code (REF): (1) 781356, (2) 782109, (3)782138. MR systems with SW version R11.1 and R12.1.Check source notice00884838098350
Not specifiedIngenia Ambition X. Product Code (REF): (1) 781356, (2) 782109, (3)782138. MR systems with SW version R11.1 and R12.1.Check source notice00884838108622
FDA code information

MR systems with SW version R11.1 and R12.1. (1) Product Code (REF): 781356; UDI-DI: 00884838090040; Serial Numbers: 47206, 47038, 47068, 47260, 47294, 47029, 47189. (2) Product Code (REF): 782109; UDI-DI: 00884838098350; Serial Numbers: 47849, 47634, 47534, 47726, 47760, 47706, 47789, 47642, 47681, 47761, 47771, 47767, 47862. (3) Product Code (REF): 782138; UDI-DI: 00884838108622; Serial Numbers: 49057.

View the original FDA recall notice

Why was this product recalled?

The potential for stiffness value errors when viewing exported MR Elastography (MRE) stiffness maps to viewer Picture Archiving and Communication System (PACS).

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Philips North America take?

Philips began notification of consignees on about 12/03/2025 via mailed Urgent Medical Device Correction letter, with a corrected notification sent on about 12/18/2025. Consignees were instructed to identify if any affected units were on hand, that the systems could continue to be use in accordance with their intended use. They were asked to provide the notification to all users of the device for awareness, maintain the Urgent Medical Device Correction letter with affected systems until the software upgrade is installed, and complete and return the provided response form. A Philips representative will contact each consignee to schedule time for a Field Service Engineer (FSE) to install a software upgrade to resolve the issues beginning in January 2026 (reference FCO78100585, FCO78100620).

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.