FDA RECALL RECORD · Z-1230-2026

PRISMAFLEX HF1000 SET, Product Code 107142; Dialyzer, High Permeability With Or Without Sealed Dialysate System Recall

PRISMAFLEX HF1000 SET, Product Code 107142; Dialyzer, High Permeability With Or Without Sealed Dialysate System

FDA recall Class II

Ongoing

FDA lists this Class II recall for PRISMAFLEX HF1000 SET, Product Code 107142; Dialyzer, High Permeability With Or Without Sealed Dialysate System. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1230-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 6, 2026
FDA report date
Feb 11, 2026
Quantity in commerce
115,112 units
Distribution
See key details
FDA cause
Component design/selection

What is being recalled?

PRISMAFLEX HF1000 SET, Product Code 107142; Dialyzer, High Permeability With Or Without Sealed Dialysate System

Affected model & UDI checker

Search the exact identifiers published for Z-1230-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedPRISMAFLEX HF1000 SET, Product Code 107142; Dialyzer, High Permeability With Or Without Sealed Dialysate SystemCheck source notice07332414069315
FDA code information

UDI/DI 07332414069315, All lots including and manufactured after 24F0094CA

View the original FDA recall notice

Why was this product recalled?

There is a potential dislodgement of the Prismaflex Set deaeration chamber from the Prismaflex Control Unit.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Vantive US Healthcare, LLC take?

Vantive issued an Urgent Medical Device Correction notice to its consignees on 01/06/2026 via USPS first class mail. The notice explained the issue, hazard involved, and requested the following: 1. You may continue to perform therapy using the Prismaflex sets listed above. Please monitor the deaeration chamber during therapy to ensure it remains in an upright position within the holder. 2. Vantive is aware of customers who have experienced this issue and attempted to further secure the deaeration chamber in an upright position. If attempting to secure the deaeration chamber in an upright position, please consider the following: . Ensure there are no kinks in the tubing and that the deaeration chamber remains visible. . If the Prismaflex Control Unit issues an Air in Blood alarm, check if air is present. If there is no air present, check if the deaeration chamber is dislodged from the holder and ensure there are no clots present. 1. If no dislodgement, please follow the operator s manual instructions for Air in Blood alarm. 2. In case of dislodgement with clotting, follow the associated instructions within the Prismaflex Control Unit operator s manual to stop the therapy and replace the set. 3. Only in case of dislodgement without clotting or any other alarms, if the chamber is secured in an upright position, follow the associated instructions within the Prismaflex Control Unit operator s manual to proceed with treatment. Facilities who further distributed the products were directed to notify those departments or facilities. Dealers, wholesalers, distributor/resellers, or original equipment manufacturers (OEM) who further distributed the products were directed to notify their customers. For general questions regarding this communication, please contact Vantive Customer Care at 833-542-CRRT (2778). Vantive Customer Care is available 24 hours a day, seven days a week.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.