FDA RECALL RECORD · Z-1233-2025

DxC 500 AU Clinical Chemistry Analyzer, REF: C63519, and C63520 with affected software. Recall

DxC 500 AU Clinical Chemistry Analyzer, REF: C63519, and C63520 with affected software.

FDA recall Class II

Ongoing

FDA lists this Class II recall for DxC 500 AU Clinical Chemistry Analyzer, REF: C63519, and C63520 with affected software.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1233-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 10, 2025
FDA report date
Mar 5, 2025
Quantity in commerce
56
Distribution
See key details
FDA cause
Software change control

What is being recalled?

DxC 500 AU Clinical Chemistry Analyzer, REF: C63519, and C63520 with affected software.

Affected model & UDI checker

Search the exact identifiers published for Z-1233-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
C63519DxC 500 AU Clinical Chemistry Analyzer, REF: C63519, and C63520 with affected software.Check source noticeNot specified
FDA code information

Software Versions: V1.4, V1.4.1, V1.4.2, V1.4.3 REF/UDI-DI/Serial Numbers: C63519/14987666545058/2024090170; C63520/14987666545065/2024070116, 2023070009, 2024010038, 2024010035, 2024050072, 2024040056, 2024030049, 2024020046, 2024070111, 2024030051, 2024030052, 2024010033, 2023090015, 2024060103, 2024060104, 2024060101, 2024080145, 2024020045, 2024020042, 2024020043, 2024020047, 2024060096, 2024040059, 2023100019, 2024030053, 2023070010, 2024070129, 2024070130, 2024020041, 2024090183, 2024010036, 2024010037, 2024070132, 2024070131, 2023100020, 2024050064, 2024050070, 2023090016, 2023110022, 2024070110, 2024060075, 2024060106, 2023110023, 2024010024, 2024010025, 2024050066, 2024010029, 2024060107, 2023080012, 2024040060, 2024050074, 2024040062, 2024010031, 2024030055, 2024080144

View the original FDA recall notice

Why was this product recalled?

Beckman Coulter has identified an issue which prevents the DxC 500 AU instrument from providing the correct onboard stability (Open Expiration) dates for assays that were loaded prior to a software upgrade (e.g. 1.3 to 1.4, 1.4 to 1.4.1, etc.). As a result, assays that have expired may still be used for sample processing, leading to inaccurate or erroneous patient test results. The issue only happens on assays loaded before instruments doing upgrades and does not impact new installation. It does not impact the assays loaded after upgrade.

What actions did BECKMAN COULTER LABORATORY SYSTEMS (SUZHOU) CO., LTD. take?

On 1/10/2025, recall notices were mailed to customers who were asked to do the following: 1) Replace and discard any reagent bottles that were on-board the analyzer at the time of any software upgrade. 2) Review Quality Control data and actions taken, for shifts out of allowed laboratory range which could have indicated potential deterioration in reagent stability. 3) Firm recommends sharing the content of this letter with your laboratory and/or Medical Director to determine if a review of previous patient test results should be conducted. 4) If specific software upgrade date is required, contact firm. 5) Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter. This issue will not occur with future software updates. If you have any questions regarding this notice, please contact the firm's Customer Support Center: Website: http://www.beckmancoulter.com Hotline: (800) 854-3633

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology