Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1243-2026

Focalyx Fusion Recall

Focalyx Fusion

FDA recall Class II

Ongoing

FDA lists this Class II recall for Focalyx Fusion. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1243-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 23, 2025
FDA report date
Feb 11, 2026
Quantity in commerce
18
Distribution
See key details
FDA cause
Software change control

What is being recalled?

Focalyx Fusion

Affected model & UDI checker

Search the exact identifiers published for Z-1243-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedFocalyx FusionCheck source notice00860003552001
FDA code information

UDI-DI: 00860003552001. Software Versions: 1.3, 1.5, 1.6. Serial Numbers: IF-FF-B-1-22080504, IF-FF-B-1-TPAA998118, IF-FF-B-1-351238, IF-FF-B-1-TPAB081335, IF-FF-B-1-22080508, IF-FF-B-1-TPAB200331, IF-FF-B-1-TPAB200333, IF-FF-B-1-TPAC505383, IF-FF-B-1-TPAB200348, IF-FF-B-1-TPAB540364, IF-FF-B-1-S5ARQL009253NYZ, IF-FF-B-1-SCARQL0014096YZ, IF-FF-B-1-TPAB532163, IF-FF-B-1-SCARQL00136033KK, IF-FF-B-1-TPAB540359, IF-FF-B-1-TPAB081326, IF-FF-B-1-22080505, IF-FF-B-1-TPAB081324

View the original FDA recall notice

Why was this product recalled?

Software device that is an accessory for image-guided interventional and diagnostic procedures involving prostate has accessories that may not function properly with Windows 10, which may cause patient harm during biopsy or cause treatment/therapy delays, so firm is recommending stop use until device is verified/validated with Windows 11.

What actions did Focalyx Technologies, LLC. take?

On 12/23/2025, correction notices started being emailed to customers informing them of the following: You will be contacted to schedule installation of the new version. An IFU will be provided at time of installation. Complete and return the response form via email to [email protected] If there are any questions, contact the firm at: focalyxtechnologies.com/contact, [email protected], (888) 636 2259, (Sel. 3 Technical Support)

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology