Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF: 80238; S 8 F x 110 cm, REF: 80239 Recall
Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF: 80238; S 8 F x 110 cm, REF: 80239
FDA recall Class I
Ongoing
FDA lists this Class I recall for Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF: 80238; S 8 F x 110 cm, REF: 80239. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1244-2025
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Feb 5, 2025
- FDA report date
- Mar 12, 2025
- Quantity in commerce
- 15,755
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF: 80238; S 8 F x 110 cm, REF: 80239
Affected model & UDI checker
Search the exact identifiers published for Z-1244-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 80236 | Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF: 80238; S 8 F x 110 cm, REF: 80239 | Check source notice | Not specified |
| 80237 | Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF: 80238; S 8 F x 110 cm, REF: 80239 | Check source notice | Not specified |
| 80238 | Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF: 80238; S 8 F x 110 cm, REF: 80239 | Check source notice | Not specified |
| 80239 | Rotarex Atherectomy System and Instructions for Use, BD SET Rotarex: S 6 F x 110 cm, REF: 80236; S 6 F x 135 cm, REF: 80237; S 8 F x 85 cm, REF: 80238; S 8 F x 110 cm, REF: 80239 | Check source notice | Not specified |
FDA code information
Instructions for Use versions prior to ZE10895 revision C1 09/24. REF/UDI-DI/Lot(Expiration): 80236/07640142811855/230893(2025-12-15), 231100(2026-06-09), 231172(2026-06-09), 231171(2026-06-09), 231526(2026-08-21), 231875(2026-08-21), 240018(2026-11-22), 240017(2026-11-22), 240375(2027-01-18), 240567(2027-01-18), 241075(2024-05-22), 241335(2024-06-19), 241550(2024-07-19), 241777(2024-08-19), 242691(2024-12-13); 80237/07640142811862/230897(2026-04-11), 230898(2026-04-11), 231033(2026-05-31), 231101(2026-06-13), 231170(2026-06-13), 231229(2026-07-07), 231400(2026-07-14), 231414(2026-07-31), 231412(2026-06-30), 231413(2026-07-31), 231411(2026-08-03), 231521(2026-09-01), 231520(2026-07-31), 231522(2026-08-31), 231523(2026-09-01), 231524(2026-09-01), 231525(2026-09-01), 231715(2026-09-22), 231885(2026-09-22), 231884(2026-10-27), 240020(2026-11-22), 240396(2026-12-08), 240530(2027-01-10), 240531(2027-01-10), 240566(2027-01-10), 240568(2027-01-25), 240754(2027-02-06), 240769(2027-02-09), 240795(2027-03-04), 241074(2027-03-07), 241119(2027-03-21), 241190(2027-04-16), 241325(2027-04-16), 241328(2027-04-25), 241336(2027-04-29), 241464(2027-05-07), 241471(2027-05-27), 241553(2027-06-05), 242019(2027-06-28), 242023(2027-06-12), 242057(2027-06-20), 242056(2027-06-20), 242112(2027-06-26), 242160(2027-07-25), 242321(2027-07-08), 242493(2027-08-01), 242596(2027-09-20), 242693(2027-11-06); 80238/07640142811879/230895(2025-11-23), 240128(2026-05-31), 240276(2026-05-31), 240532(2026-05-31), 240658(2026-09-30), 241076(2026-09-30), 241330(2026-09-30), 241548(2026-09-30), 241768(2026-09-30), 241911(2026-09-30), 242117(2026-09-30), 242494(2026-09-30), 242521(2026-09-30), 242519(2026-05-31), 242520(2026-09-30); 80239/07640142811886/230896(2025-11-23), 231231(2026-06-13), 231410(2026-06-13), 231527(2026-08-31), 231883(2026-10-19), 232029(2026-08-31), 232028(2026-10-19), 240129(2026-11-30), 240278(2026-11-30), 240557(2027-01-22), 240746(2027-01-22), 240794(2027-02-28), 241077(2027-04-08), 241188(2027-04-08), 241462(2027-05-03), 241523(2027-05-03), 241767(2027-06-26), 241910(2027-06-26), 242159(2027-08-09), 242322(2027-09-10), 242518(2027-09-10), 242694(2027-10-25), 242695(2027-10-25)
Why was this product recalled?
Atherectomy Catheter eIFU updated to clarify and emphasize procedural steps intended to reduce the likelihood of catheter breakage events. Catheter has an outer cylinder connected to a rotating helix, which could fracture and/or break, which would require retrieval, and helix facture/break could cause vessel injury and lead to severe bleeding.
What actions did Bard Peripheral Vascular, Inc. take?
On 2/5/2025, "URGENT: MEDICAL DEVICE CORRECTION" notices were sent to customers informing them of the following: 1) In order to minimize risks of illness or injury, please review the updates to the eIFU in their entirety. The updated eIFU can be found on BD's website at https://www.bardpv.com/Rotarex-IFU.php 2) Firm recommends that end users review the updated on-line, on-demand training on the safe and effective use of the Rotarex Atherectomy System, reflecting the recent eIFU updates. The online education may be completed by registering on the BDLA Learning Management System https://academy.bd.com. 3) If possible, post this notice with the stored product as evidence of the updated eIFU. 4) Share this notification with all users within your facility network of the product to ensure they are also aware of the updated eIFU. This correction notice should be shared with anyone who needs to be aware within your organization and forwarded to any organization where potentially affected devices have been transferred. 5) Complete and return the response form via email to [email protected] Report any complaints experienced with the use of this product to the firm's Regional Complaint Center: 1-844-8BD-LIFE (1-844-823-5433) Say "Product Complaints" when prompted Mon-Fri 8:00am to 5:00pm CT Email: [email protected] For additional information regarding Advance Clinical Education and Training Programs, please email the Advance Team at [email protected] Distributors were provided with a copy of the customer communication (correction notice) and asked to provide a copy to all customers on BD's behalf. If you have any questions, call 480-303-2602.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1244-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology