Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1253-2025

Baxter Sigma Spectrum Infusion System (V6 Platform), Product Code 35700BAX Recall

Baxter Sigma Spectrum Infusion System (V6 Platform), Product Code 35700BAX

FDA recall Class I

Ongoing

FDA lists this Class I recall for Baxter Sigma Spectrum Infusion System (V6 Platform), Product Code 35700BAX. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1253-2025
Recall classification
Class I
Enforcement status
Ongoing
Initiation date
Feb 5, 2025
FDA report date
Mar 12, 2025
Quantity in commerce
66 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Baxter Sigma Spectrum Infusion System (V6 Platform), Product Code 35700BAX

Affected model & UDI checker

Search the exact identifiers published for Z-1253-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedBaxter Sigma Spectrum Infusion System (V6 Platform), Product Code 35700BAXCheck source notice00085412091570
FDA code information

UDI/DI 00085412091570, Serial Numbers: 704198, 728888, 751492, 758399, 758614, 761662, 763197, 766291, 772611, 773302, 775486, 778837, 785425, 793453, 795240, 798065, 814485, 818850, 826637, 850214, 850854, 851500, 854157, 855803, 863354, 870307, 873114, 876483, 877014, 877102, 889153, 896292, 900181, 906422, 907176, 912242, 913562, 927864, 938265, 958176, 962828, 963975, 965158, 974676, 981605, 982459, 984154, 986132, 986634, 989053, 993087, 994417, 994842, 995538, 997153, 1006359, 1011058, 1013925, 1014064, 1014529, 1014856, 1017150, 1017546, 1018229, 1018703, 1024109

View the original FDA recall notice

Why was this product recalled?

There is the potential for missing motor mounting screws, which may have occurred during the servicing process.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Baxter Healthcare Corporation take?

Baxter issued an "URGENT MEDICAL DEVICE RECAL" notice to its consignees on 02/05/2025 via US Mail. The notice explained the issue, hazard involved, and requested the following actions be taken: "1. Immediately locate, isolate, and cease all use of Spectrum pumps with the affected serial numbers. The product code and serial number can be found on the bottom of the infusion pump. 2. Contact Baxter to arrange for the return of the affected pumps for inspection and reservicing as applicable. 3. If you received this communication directly from Baxter, acknowledge receipt on the enclosed reply form even if you have no remaining inventory. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 4. If you purchased this product from a distributor or wholesaler, please note that responding on the Baxter customer portal is not applicable. If a response is requested by your distributor or wholesaler, please respond to them according to their instructions. 5. Please forward a copy of this communication to all departments within your institution who use the affected product. 6. If you are a dealer, wholesaler, distributor/reseller, or original equipment manufacturer (OEM) that distributed any affected product to other facilities, please conduct a user-level recall of the affected product that you distributed to customers and check the associated box on the customer portal. For general questions regarding this communication, or quality problems, please contact your Baxter sales representative, or Baxter Global Technical Services at 800-843-7867 Monday through Friday, between 7:00 am and 7:00 pm Eastern Time."

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology