Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1255-2026

Vortex Surgical 25GA Backflush, VS0270.25; 25GA Backflush Retractable, VS0275.27 Recall

Vortex Surgical 25GA Backflush, VS0270.25; 25GA Backflush Retractable, VS0275.27

FDA recall Class II

Ongoing

FDA lists this Class II recall for Vortex Surgical 25GA Backflush, VS0270.25; 25GA Backflush Retractable, VS0275.27. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1255-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 16, 2025
FDA report date
Feb 11, 2026
Quantity in commerce
14,789 (8651 US; 6138 OUS)
Distribution
See key details
FDA cause
Employee error

What is being recalled?

Vortex Surgical 25GA Backflush, VS0270.25; 25GA Backflush Retractable, VS0275.27

Affected model & UDI checker

Search the exact identifiers published for Z-1255-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedVortex Surgical 25GA Backflush, VS0270.25; 25GA Backflush Retractable, VS0275.27Check source notice810123480470
Not specifiedVortex Surgical 25GA Backflush, VS0270.25; 25GA Backflush Retractable, VS0275.27Check source notice810123480753
Not specifiedVortex Surgical 25GA Backflush, VS0270.25; 25GA Backflush Retractable, VS0275.27Check source notice810123480517
Not specifiedVortex Surgical 25GA Backflush, VS0270.25; 25GA Backflush Retractable, VS0275.27Check source notice810123480791
FDA code information

25GA Backflush Pouch UDI 810123480470 Box UDI 810123480753 Lot 2411017; 27GA Backflush Retractable Pouch UDI 810123480517 Box UDI 810123480791 Lot 2411016

View the original FDA recall notice

Why was this product recalled?

There may be voids located in the seal of Tyvek pouches associated with identified catalog numbers and lots. A compromised sterile barrier has potential of bioburden contamination which could lead to infection.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Vortex Surgical Inc. take?

On December 16, 2025, the firm began notifying customers via "URGENT: MEDICAL DEVICE RECALL" letters. Customers were instructed to discontinue use of affected product and to return product for Vortex Surgical to evaluate the seals; use UPS account 23623Y to ship unused products to Vortex Surgical Inc for replacement or refund at: Attn: Andrew Ritts Recall Tyvek Seal, Vortex Surgical Inc., 4 Research Park Dr., Suite 124, St. Charles, MO 63304; complete and return the Acknowledgement and Receipt Form via mail or email to: [email protected]. Actions have been implemented to improve training of individuals associated with sealing and inspecting seals to prevent reoccurrence. The sealer associated with the issue is being updated to include a guard to prevent reoccurrence. If you have any questions, contact the Director of Regulatory Affairs at 635-778-4350 or email to: [email protected] from 8am to 5pm.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.