FDA RECALL RECORD · Z-1267-2025

MIM software; System, Image Processing, Radiological Recall

MIM software; System, Image Processing, Radiological

FDA recall Class II

Ongoing

FDA lists this Class II recall for MIM software; System, Image Processing, Radiological. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1267-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 19, 2025
FDA report date
Mar 12, 2025
Quantity in commerce
352 units
Distribution
See key details
FDA cause
Software design

What is being recalled?

MIM software; System, Image Processing, Radiological

Affected model & UDI checker

Search the exact identifiers published for Z-1267-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMIM software; System, Image Processing, RadiologicalCheck source notice00850009343192
FDA code information

Software versions: 1. MIM Anyware 1.1, UDI/DI Not Applicable 2. MIM 7.2.0, UDI/DI 00850009343192; 3. MIM 7.2.1, UDI/DI 00850009343192; 4. MIM 7.2.2, UDI/DI 00850009343192; 5. MIM 7.2.3, UDI/DI 00850009343192; 6. MIM 7.2.4, UDI/DI 00850009343192; 7. MIM 7.2.5, UDI/DI 00850009343192; 8. MIM 7.2.6, UDI/DI 00850009343192

View the original FDA recall notice

Why was this product recalled?

In situations where two images with differing Fields of View (FOV) complete an image fusion, an incorrect, elevated Maximum Standardized Uptake Value (SUV) within MIM Software versions 7.2.0 through 7.2.6.could result.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Mim Software, Inc. take?

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/19/2025 disseminated using a traceable means. The notice explained the issue, potential risk to the patient, and provided the following actions to be taken to use the device: "You can continue to use your device if you are not using the SUV tool or if you access a non-impacted version of MIM when using the SUV tool. If using a non-default workflow and fusing a PET or SPECT image with another image with a different FOV, do not incorporate the Maximum SUV value into assessment of disease status, progression or response to treatment. NOTE: MIM s default workflows do not fuse images with differing FOV, and the Maximum SUV values reported are correct. If you are using the SUV tool and are unsure if you are affected by this issue, please contact your MIM Site Development Manager, your Distributor for MIM Software, or MIM Customer Support." "Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions."

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology