DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement Recall
DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement
FDA recall Class II
Ongoing
FDA lists this Class II recall for DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1291-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jan 31, 2025
- FDA report date
- Mar 12, 2025
- Quantity in commerce
- 714 cases
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement
Affected model & UDI checker
Search the exact identifiers published for Z-1291-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 71-C5100NS | DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement | Check source notice | Not specified |
| 71-C6100NS | DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement | Check source notice | Not specified |
| 71-C7100NS | DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement | Check source notice | Not specified |
| Not specified | DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement | Check source notice | 00749756578738 |
| Not specified | DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement | Check source notice | 00749756578851 |
| Not specified | DeRoyal REF 71-C5100NS Tubing, Suction, Connector; DeRoyal REF 71-C6100NS Tubing, Suction, Connector; DeRoyal REF 71-C7100NS Tubing, Suction, Connector Indicated use - Fluid Movement | Check source notice | 00749756578967 |
FDA code information
71-C5100NS Bulk Roll Suction Connector Tubing UDI-DI code: 00749756578738 Lot Number: CNWKD04-04 71-C6100NS Bulk Roll Suction Connector Tubing UDI-DI code: 00749756578851 Lot Numbers: CNWKD04-04 CNWKD04-05 CNWKD04-10 CNWKD09-08 71-C7100NS Bulk Roll Suction Connector Tubing UDI-DI code: 00749756578967 Lot Numbers: CNWKD04-06 CNWKD04-07
Why was this product recalled?
Due to the inner diameter of the suction tubing connector being smaller than the suction canister lid port connection. This size difference, combined with the durometer of the plastic material, makes the assembly to a suction canister lid port difficult. This could result in inadequate suction and cause a delay in patient care.
These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.
What actions did DEROYAL INDUSTRIES, INC. take?
On 01/31/2025, the firm sent an "URGENT: MEDICAL DEVICE RECALL" Letter via email to customers informing them that bulk roll, non-sterile suction connector tubing products and lots which could possibly be defective. The inner diameter of the suction tubing connector, is smaller than the suction canister lid port connection. This size difference, combined with the durometer of the plastic material, makes the assembly to a suction canister lid port difficult. This could result in inadequate suction and cause a delay in patient care. Customers are instructed to: 1) Identify affected product numbers in their inventory and place in quarantine to prevent further use. Destroy all product identified and complete the NOTICE OF RETURN FORM indicating all affected products found in their inventory. Return the completed form to DeRoyal via fax to 865-362-3716 or [email protected] no later than March 7, 2025. Complete and return the Notice of Return Form even if you no longer have any invenorty. 2) If affected products have been further distributed to other facilities or departments within customer's institution, customers should be notified of this recall. by either directly forwarding the recall notification and notify DeRoyal that you have done so, or you may provide us with the customer listing and we will contact them. 3) Delay in patient care could lead to serious adverse events. A patient may aspirate, requiring immediate suction to prevent life-threatening injury. Inadequate suction not identified by the clinician could increase infection risk. For questions regarding credit for the recall, contact Jennifer Marsh at [email protected] or by phone 865-362-6206 or email us at [email protected]. If customers prefer to receive replacement product, contact DeRoyal Sales Representative or contact Angie Schubert, Assistant Product Director, at [email protected].
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1291-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 20, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 20, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology