Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1297-2025

GE Healthcare Revolution Apex, System, X-ray, Tomography, Computed Recall

GE Healthcare Revolution Apex, System, X-ray, Tomography, Computed

FDA recall Class II

Ongoing

FDA lists this Class II recall for GE Healthcare Revolution Apex, System, X-ray, Tomography, Computed. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1297-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 18, 2025
FDA report date
Mar 19, 2025
Quantity in commerce
56 units
Distribution
See key details
FDA cause
Process change control

What is being recalled?

GE Healthcare Revolution Apex, System, X-ray, Tomography, Computed

Affected model & UDI checker

Search the exact identifiers published for Z-1297-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedGE Healthcare Revolution Apex, System, X-ray, Tomography, ComputedCheck source notice00195278460271
Not specifiedGE Healthcare Revolution Apex, System, X-ray, Tomography, ComputedCheck source notice00840682118552
Not specifiedGE Healthcare Revolution Apex, System, X-ray, Tomography, ComputedCheck source notice00840682146616
Not specifiedGE Healthcare Revolution Apex, System, X-ray, Tomography, ComputedCheck source notice00195278469700
Not specifiedGE Healthcare Revolution Apex, System, X-ray, Tomography, ComputedCheck source notice00840682123471
Not specifiedGE Healthcare Revolution Apex, System, X-ray, Tomography, ComputedCheck source notice00195278633149
FDA code information

System ID Number: 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN; System ID Number: 650493REV1, UDI To be provided, Mfg. Lot/Serial Number REVVX1900035CN; System ID Number: 630266CTRV, UDI (01)00195278460271(11)240800(21)REV2X2400093CN, Mfg. Lot/Serial Number REV2X2400093CN; System ID Number: 773702REVCT6ED, UDI (01)00195278460271(11)231000(21)REV2X2300071CN, Mfg. Lot/Serial Number REV2X2300071CN; System ID Number: 312996CTRV2, UDI (01)00195278460271(11)241000(21)REV2X2400126CN, Mfg. Lot/Serial Number REV2X2400126CN; System ID Number: 1000000018, UDI To be provided, Mfg. Lot/Serial Number REV2X2300074CN; System ID Number: 617726Y6REV2, UDI (01)00840682118552(11)170900(21)REVVX1700090CN, Mfg. Lot/Serial Number REVVX1700090CN; System ID Number: 617726Y6REV1, UDI (01)00840682118552(11)160700(21)REVVX1600018CN, Mfg. Lot/Serial Number REVVX1600018CN; System ID Number: 336890APEX, UDI (01)00195278460271(11)240900(21)REV2X2400116CN, Mfg. Lot/Serial Number REV2X2400116CN; System ID Number: 704801APEX, UDI (01)00195278460271(11)241000(21)REV2X2400139CN, Mfg. Lot/Serial Number REV2X2400139CN; System ID Number: 336832APEX, UDI (01)00195278460271(11)241000(21)REV2X2400129CN, Mfg. Lot/Serial Number REV2X2400129CN; System ID Number: 980993CT1, UDI (01)00840682146616(11)210700(21)REV2A2100029CN, Mfg. Lot/Serial Number REV2A2100029CN; System ID Number: 910715REV1, UDI (01)00195278460271(11)241200(21)REV2X2400155CN, Mfg. Lot/Serial Number REV2X2400155CN; System ID Number: 603650APEX1, UDI (01)00195278460271(11)240600(21)REV2X2400068CN, Mfg. Lot/Serial Number REV2X2400068CN; System ID Number: 603526APEX, UDI (01)00195278469700(11)230100(21)REV242200004CN, Mfg. Lot/Serial Number REV242200004CN; System ID Number: 973971REV161, UDI (01)00195278460271(11)241200(21)REV2X2400158CN, Mfg. Lot/Serial Number REV2X2400158CN; System ID Number: 914787APEX, UDI (01)00195278460271(11)231000(21)REV2X2300081CN, Mfg. Lot/Serial Number REV2X2300081CN; System ID Number: 718963EDCT, UDI (01)00195278460271(11)240600(21)REV2X2400070CN, Mfg. Lot/Serial Number REV2X2400070CN; System ID Number: 315473APEX1, UDI (01)00195278460271(11)241100(21)REV2X2400152CN, Mfg. Lot/Serial Number REV2X2400152CN; System ID Number: 614RMHAPEX1, UDI (01)00195278460271(11)240300(21)REV2X2400026CN, Mfg. Lot/Serial Number REV2X2400026CN; System ID Number: 724357APEX, UDI (01)00195278460271(11)230800(21)REV2X2300046CN, Mfg. Lot/Serial Number REV2X2300046CN; System ID Number: 843797REVCT2, UDI (01)00840682123471(11)211100(21)REVV82100041CN, Mfg. Lot/Serial Number REVV82100041CN; System ID Number: 806775APEX, UDI (01)00195278460271(11)231000(21)REV2X2300075CN, Mfg. Lot/Serial Number REV2X2300075CN; System ID Number: 703776APEX, UDI To be provided, Mfg. Lot/Serial Number REV2X2400167CN; System ID Number: 414805REVCT, UDI Not Applicable, Mfg. Lot/Serial Number 00000441066CN8; System ID Number: 262577REVCT, UDI (01)00195278460271(11)240500(21)REV2X2400055CN, Mfg. Lot/Serial Number REV2X2400055CN; System ID Number: 832CT000059, UDI (01)00840682146616(11)220100(21)REV2A2200006CN, Mfg. Lot/Serial Number REV2A2200006CN; System ID Number: 833CT000155, UDI (01)00195278460271(11)240500(21)REV2X2400053CN, Mfg. Lot/Serial Number REV2X2400053CN; System ID Number: AH7084CT02, UDI (01)00840682146616(11)220100(21)REV2A2200007CN, Mfg. Lot/Serial Number REV2A2200007CN; System ID Number: AH1923CT01, UDI (01)00195278460271(11)240300(21)REV2X2400024CN, Mfg. Lot/Serial Number REV2X2400024CN; System ID Number: BG4549CT04, UDI (01)00195278460271(11)230600(21)REV2X2300026CN, Mfg. Lot/Serial Number REV2X2300026CN; System ID Number: E416249APEXCTE, UDI (01)00195278460271(11)231100(21)REV2X2300087CN, Mfg. Lot/Serial Number REV2X2300087CN; System ID Number: 082421140968, UDI To be provided, Mfg. Lot/Serial Number Not Available; System ID Number: 082421140960, UDI (01)00195278633149(11)240500(21)CBFCG2400006HM, Mfg. Lot/Serial Number CBFCG2400006HM; System ID Number: 082421140959, UDI (01)00195278633149(11)240500(21)CBFCG2400004HM, Mfg. Lot/Serial Number CBFCG2400004HM; System ID Number: DK1044CT12, UDI (01)00840682146616(11)200200(21)REV2A2000001CN, Mfg. Lot/Serial Number REV2A2000001CN; System ID Number: DK1547CT01, UDI (01)00840682123471(11)210100(21)REVV82100002CN, Mfg. Lot/Serial Number REVV82100002CN; System ID Number: DK1051CT04, UDI To be provided, Mfg. Lot/Serial Number 00000442412CN3; System ID Number: CT52261701, UDI To be provided, Mfg. Lot/Serial Number REV2X2300050CN; System ID Number: M40230104, UDI (01)00840682118552(11)160600(21)REVVX1600012CN, Mfg. Lot/Serial Number REVVX1600012CN; System ID Number: REV020, UDI To be provided, Mfg. Lot/Serial Number 00000442334CN9; System ID Number: CT00486, UDI (01)00195278460271(11)231000(21)REV2X2300079CN, Mfg. Lot/Serial Number REV2X2300079CN; System ID Number: KRCT54972221, UDI To be provided, Mfg. Lot/Serial Number Not Available; System ID Number: KRCT54893311, UDI To be provided, Mfg. Lot/Serial Number Not Available; System ID Number: 0850211084, UDI To be provided, Mfg. Lot/Serial Number REV2X2300016CN; System ID Number: NO1068CT04, UDI (01)00840682118552(11)171000(21)REVVX1700106CN, Mfg. Lot/Serial Number REVVX1700106CN; System ID Number: P00652CT01, UDI (01)00195278460271(11)240200(21)REV2X2400017CN, Mfg. Lot/Serial Number REV2X2400017CN; System ID Number: PL1672CT03, UDI (01)00840682146616(11)210600(21)REV2A2100019CN, Mfg. Lot/Serial Number REV2A2100019CN; System ID Number: SA1420CT01, UDI To be provided, Mfg. Lot/Serial Number REV2X2400137CN; System ID Number: SA2973CT01, UDI To be provided, Mfg. Lot/Serial Number REV2X23000030CN; System ID Number: CT53972381, UDI To be provided, Mfg. Lot/Serial Number REV2X2400121CN; System ID Number: A201CT07, UDI (01)00195278460271(11)240200(21)REV2X2400014CN, Mfg. Lot/Serial Number REV2X2400014CN; System ID Number: AC001CT05, UDI (01)00840682118552(11)181000(21)REVVX1800101CN, Mfg. Lot/Serial Number REVVX1800101CN; System ID Number: 541010CT02, UDI (01)00840682118552(11)190900(21)REVVX1900106CN, Mfg. Lot/Serial Number REVVX1900106CN; System ID Number: THCTX24524601, UDI (01)00195278460271(11)241000(21)REV2X2400131CN, Mfg. Lot/Serial Number REV2X2400131CN; System ID Number: AE1025CT03, UDI (01)00840682118552(11)210300(21)REVVX2100010CN, Mfg. Lot/Serial Number REVVX2100010CN.

View the original FDA recall notice

Why was this product recalled?

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did GE Medical Systems, LLC take?

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/18/2025 via letter delivered using a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken: "You can continue to use your system. Please monitor the floor near the base of the gantry for any accumulation of blue fluid. If blue fluid is observed, wear nitrile or rubber gloves, clean up the fluid by absorbing it with a noncombustible absorbent material, and dispose of the absorbent material in a suitable container for disposal. Please contact your local GE HealthCare service representative if any blue fluid is observed. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please complete and return the attached acknowledgement form to [email protected]."

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology