FDA RECALL RECORD · Z-1301-2025

GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed Recall

GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed

FDA recall Class II

Ongoing

FDA lists this Class II recall for GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1301-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 18, 2025
FDA report date
Mar 19, 2025
Quantity in commerce
1 unit
Distribution
See key details
FDA cause
Process change control

What is being recalled?

GE Healthcare Revolution Apex Select, System, X-ray, Tomography, Computed

Affected model & UDI checker

Search the exact identifiers published for Z-1301-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedGE Healthcare Revolution Apex Select, System, X-ray, Tomography, ComputedCheck source notice00195278460271
FDA code information

System ID Number 907729APEX1, UDI (01)00195278460271(11)231100(21)REV2X2300082CN, Mfg. Lot/Serial Number REV2X2300082CN

View the original FDA recall notice

Why was this product recalled?

GE HealthCare has become aware of a potential issue on certain Revolution CT and Revolution Apex series CT systems (please see affected product details and the Appendix) that can result in coolant fluid (glycol) leaking onto the floor near the base of the gantry. The fluid is visible as it has a blue color. It is not corrosive and is not at a harmful temperature. However, there is a potential for injury if the fluid is not noticed and a person slips and/or falls.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did GE Medical Systems, LLC take?

GE Healthcare issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 02/18/2025 via letter delivered using a traceable means. The notice explained the issue, safety issue, and requested the following actions be taken: "You can continue to use your system. Please monitor the floor near the base of the gantry for any accumulation of blue fluid. If blue fluid is observed, wear nitrile or rubber gloves, clean up the fluid by absorbing it with a noncombustible absorbent material, and dispose of the absorbent material in a suitable container for disposal. Please contact your local GE HealthCare service representative if any blue fluid is observed. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please complete and return the attached acknowledgement form to [email protected]."

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology