Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1313-2025

Brand Name: TruScan Body, REF CN38-SB Product Name: TruScan Body compatible with CONELOG 3.8 Recall

Brand Name: TruScan Body, REF CN38-SB Product Name: TruScan Body compatible with CONELOG 3.8 Model/Catalog Number: CN38-SB Software Version: N/A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: TruScan Body, REF CN38-SB Product Name: TruScan Body compatible with CONELOG 3.8. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1313-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 27, 2023
FDA report date
Mar 19, 2025
Quantity in commerce
61 units
Distribution
See key details
FDA cause
Unknown/Undetermined by firm

What is being recalled?

Brand Name: TruScan Body, REF CN38-SB Product Name: TruScan Body compatible with CONELOG 3.8 Model/Catalog Number: CN38-SB Software Version: N/A

Affected model & UDI checker

Search the exact identifiers published for Z-1313-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
CN38-SBBrand Name: TruScan Body, REF CN38-SB Product Name: TruScan Body compatible with CONELOG 3.8Check source noticeNot specified
Not specifiedBrand Name: TruScan Body, REF CN38-SB Product Name: TruScan Body compatible with CONELOG 3.8Check source notice08800076411791
FDA code information

Lot Code: KCCBB1216110/ UDI: (01)08800076411791

View the original FDA recall notice

Why was this product recalled?

Retroactive: Due to defect and quality issues dental scanner may not scan properly/accurately

What actions did TruAbutment Inc. take?

On February 3, 2025, TruAbutment Inc. begin to reach out to affected customers about the recall. On February 13, 2025, TruAbutment Inc. followed up with consignees by phone and followed up the phone call with a "Urgent: Medical Device Recall" Notification. TruAbutment asked consignees to take the following actions: 1. Stop usage and quarantine all CN38-SB until your on-hand products are verified. 2. Check all unopened CN38-SB. a. Please check the Lot # and quarantine the products with Lot # KCCBB1216110. 3. Then check the opened CN38-SB products. 4. Please quarantine any unaligned products. 4. Please return the quarantined products back to TruAbutment with the signed acknowledgement and replacement form that is attached to this letter. 5.If the product was transferred or distributed to a different facility, please forward this notification and inform TruAbutment Inc. by contacting the customer support team with the information below.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.