Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheters: Livewire Steerable Electrophysiology Catheter,10 Elec. 2-5-2mm 401582RH 401940RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-10-2mm 401904RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-5-2mm 401914RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-8-2-60-2-8-2mm 401932RH; Livewire Steerable Electrophysiology Catheter,8 Elec. 2-2-2mm 401652RH; Recall
Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheters: Livewire Steerable Electrophysiology Catheter,10 Elec. 2-5-2mm 401582RH 401940RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-10-2mm 401904RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-5-2mm 401914RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-8-2-60-2-8-2mm 401932RH; Livewire Steerable Electrophysiology Catheter,8 Elec. 2-2-2mm 401652RH;
FDA recall Class I
Ongoing
FDA lists this Class I recall for Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheters: Livewire Steerable Electrophysiology Catheter,10 Elec. 2-5-2mm 401582RH 401940RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-10-2mm 401904RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-5-2mm 401914RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-8-2-60-2-8-2mm 401932RH; Livewire Steerable Electrophysiology Catheter,8 Elec. 2-2-2mm 401652RH;. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1321-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Dec 22, 2025
- FDA report date
- Feb 18, 2026
- Quantity in commerce
- 41
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheters: Livewire Steerable Electrophysiology Catheter,10 Elec. 2-5-2mm 401582RH 401940RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-10-2mm 401904RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-5-2mm 401914RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-8-2-60-2-8-2mm 401932RH; Livewire Steerable Electrophysiology Catheter,8 Elec. 2-2-2mm 401652RH;
Affected model & UDI checker
Search the exact identifiers published for Z-1321-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| Not specified | Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheters: Livewire Steerable Electrophysiology Catheter,10 Elec. 2-5-2mm 401582RH 401940RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-10-2mm 401904RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-5-2mm 401914RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-8-2-60-2-8-2mm 401932RH; Livewire Steerable Electrophysiology Catheter,8 Elec. 2-2-2mm 401652RH; | Check source notice | 10888277407480 |
| Not specified | Reprocessed Livewire Steerable Diagnostic Electrophysiology Catheters: Livewire Steerable Electrophysiology Catheter,10 Elec. 2-5-2mm 401582RH 401940RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-10-2mm 401904RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-5-2mm 401914RH; Livewire Steerable Electrophysiology Catheter,20 Elec. 2-8-2-60-2-8-2mm 401932RH; Livewire Steerable Electrophysiology Catheter,8 Elec. 2-2-2mm 401652RH; | Check source notice | 10888277407572 |
FDA code information
401582RH UDI-DI (case) 20888277407517¿ (ea) 10888277407510 Lots EP250618 EP250630; 401940RH UDI-DI 10888277407480 Lot EP250618; 401904RH UDI-DI (case) 20888277407609¿ (ea) 10888277407602 Lots EP250618 EP250630; 401914RH UDI-DI (case) 20888277407524¿ (ea) 10888277407527 Lot EP250618; 401932RH UDI-DI (case) 20888277407678¿ (ea) 10888277407671 Lots EP250618 EP250630; 401652RH UDI-DI 10888277407572 Lots EP250618 EP250630
Why was this product recalled?
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
What actions did Medline Industries, LP take?
On December 22, 2025, the firm issued Urgent Medical Device Recall letters as a follow-up to an earlier action in 2025 for the same issue, but newly identified item numbers and lots of previously identified item numbers. Notably, the products associated with the new/updated action are to be destroyed by the customer on-site for credit, and not returned to the manufacturer. ***Updated 5/18/26*** The firm issued an expanded notification on 5/15/26 which expanded the action to include on additional lot of product. These devices were not expired at the time of the original notification but have since expired.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1321-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology