Medline ReNewal Reprocessed ViewFlex Xtra ICE Catheters: ViewFlex Xtra ICE Catheter 9F x 90cm D087031RH 20197344019554 Recall
Medline ReNewal Reprocessed ViewFlex Xtra ICE Catheters: ViewFlex Xtra ICE Catheter 9F x 90cm D087031RH 20197344019554¿
FDA recall Class I
Ongoing
FDA lists this Class I recall for Medline ReNewal Reprocessed ViewFlex Xtra ICE Catheters: ViewFlex Xtra ICE Catheter 9F x 90cm D087031RH 20197344019554. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1323-2026
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Dec 22, 2025
- FDA report date
- Feb 18, 2026
- Quantity in commerce
- 650
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Medline ReNewal Reprocessed ViewFlex Xtra ICE Catheters: ViewFlex Xtra ICE Catheter 9F x 90cm D087031RH 20197344019554
The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.
Search FDA identifier text
Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.
This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.
FDA code information
UDI-DI (case) 20197344019554¿ (ea) 10197344019557 Lots EP250324 EP250416 EP250512 EP250521 EP250609 EP250630 EP250710 EP250724
Why was this product recalled?
These lots of reprocessed catheters may contain small particles of residual material. If affected devices are used, there is a risk of inflammatory response, systemic infection, or granulomatous reaction or thrombus reaction which may lead to cerebral or pulmonary embolism, or deep vein thrombosis. Expansion of recall Z-2610/2614-2025
What actions did Medline Industries, LP take?
On December 22, 2025, the firm issued Urgent Medical Device Recall letters as a follow-up to an earlier action in 2025 for the same issue, but newly identified item numbers and lots of previously identified item numbers. Notably, the products associated with the new/updated action are to be destroyed by the customer on-site for credit, and not returned to the manufacturer. ***Updated 5/18/26*** The firm issued an expanded notification on 5/15/26 which expanded the action to include on additional lot of product. These devices were not expired at the time of the original notification but have since expired.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1323-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology