Brand Name: Da Vinci 5 Product Name: ASSY,DV5 CONSOLE,IS5000 Foot Tray Pedals Recall
Brand Name: Da Vinci 5 Product Name: ASSY,DV5 CONSOLE,IS5000 Foot Tray Pedals Model/Catalog Number: 380730 Software Version: N/A . Refer to HHE for additional details. Component: N/A
FDA recall Class II
Ongoing
FDA lists this Class II recall for Brand Name: Da Vinci 5 Product Name: ASSY,DV5 CONSOLE,IS5000 Foot Tray Pedals. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1325-2025
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Feb 26, 2025
- FDA report date
- Mar 19, 2025
- Quantity in commerce
- 439 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
Brand Name: Da Vinci 5 Product Name: ASSY,DV5 CONSOLE,IS5000 Foot Tray Pedals Model/Catalog Number: 380730 Software Version: N/A . Refer to HHE for additional details. Component: N/A
Affected model & UDI checker
Search the exact identifiers published for Z-1325-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 380730 | Brand Name: Da Vinci 5 Product Name: ASSY,DV5 CONSOLE,IS5000 Foot Tray Pedals | Check source notice | Not specified |
| Not specified | Brand Name: Da Vinci 5 Product Name: ASSY,DV5 CONSOLE,IS5000 Foot Tray Pedals | Check source notice | 00886874119747 |
FDA code information
UDI: 00886874119747/ All da Vinci dV5 consoles and foot trays assemblies manufactured prior to 12/20/2024.
Why was this product recalled?
Due to an increase in complaints concerning foot tray pedal spring failing resulting in the pedal remaining pressed
What actions did Intuitive Surgical, Inc. take?
On February 26, 2025, Intuitive Surgical issued a Urgent: Medical Device Correction to affected consignees via E-Mail. Intuitive asked consignees to take the following actions: 1. Read and understand this notification. 2. Share this notification with all personnel at your site who perform and support da Vinci procedures. 3. Inform affected personnel when the correction has been completed. 4. Complete the attached Acknowledgement Form immediately and return it via fax or email to Intuitive as instructed on the form. 5. Please retain a copy of this letter and the acknowledgement form for your files. 6. Please inform Intuitive of any Adverse Events* or quality problems concerning the use of the subject devices via the standard complaint process 7. Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA's MedWatch Adverse Event Reporting program either online, by regular mail or by fax. 8. Complete and submit the report Online Regular Mail or Fax: Download form or call 1- 800-332-1088 to request a reporting form, then complete and return to the address on the pre-addressed form or submit by fax to 1-800-FDA-0178. An Intuitive representative will schedule a visit to replace the affected foot tray assembly.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1325-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 22, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 22, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology