Medline medical convenience kits packaged as: ADULT CATH SECUREMENT DRSG, Kit SKU DYNDC3109 Recall

Medline medical convenience kits packaged as: ADULT CATH SECUREMENT DRSG, Kit SKU DYNDC3109

FDA recall Class II

Ongoing

FDA lists this Class II recall for Medline medical convenience kits packaged as: ADULT CATH SECUREMENT DRSG, Kit SKU DYNDC3109. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1327-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 24, 2025
FDA report date
Feb 18, 2026
Quantity in commerce
1350 kits
Distribution
See key details
FDA cause
Packaging change control

What is being recalled?

Medline medical convenience kits packaged as: ADULT CATH SECUREMENT DRSG, Kit SKU DYNDC3109

Affected model & UDI checker

Search the exact identifiers published for Z-1327-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
Not specifiedMedline medical convenience kits packaged as: ADULT CATH SECUREMENT DRSG, Kit SKU DYNDC3109Check source notice10193489918113
FDA code information

UDI/DI 10193489918113, Lot Number 25EBD613

View the original FDA recall notice

Why was this product recalled?

Medline Industries, LP is issuing a recall for specific items and lots of Medline Kits containing BD ChloraPrep Triple Swabsticks which BD determined may exhibit an open seal on the packaging of the applicators.

What actions did Medline Industries, LP take?

Medline issued an URGENT MEDICAL DEVICE recall notice to its consignees on 12/24/2025 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the consignee locate and quarantine the affected units and respond to the notice using the website link: https://recalls.medline.com; Recall Reference #: R-25-257-FGX1, and the specific code listed in the notice. Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable. Distributors and those who further distributed the affected its were directed to notify those to whom the kits were distributed or transferred, and include those quantities on the response form. Upon receipt of your submitted response form, you will receive over-labels to provide to your customers to affix to their affected product. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.