FDA lists this Class II recall for Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA;. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1341-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 13, 2026
FDA report date
Feb 18, 2026
Quantity in commerce
1202 units (537 US, 665 OUS) (2/23/2026: 574 US, 1239 units)
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; 3) LCSU 4, RTCA; Catalog Number: 881152;

Affected model & UDI checker

Search the exact identifiers published for Z-1341-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
880052Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA;Check source noticeNot specified
880062Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA;Check source noticeNot specified
881152Laerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA;Check source noticeNot specified
Not specifiedLaerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA;Check source notice07045432088519
Not specifiedLaerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA;Check source notice07045432088533
Not specifiedLaerdal Compact Suction Unit 4 RTCA version (LCSU 4 RTCA) 1) LCSU 4, 800 ml, RTCA;Check source notice07045432067132
FDA code information

1) LCSU 4, 800 ml, RTCA; Catalog Number: 880052; UDI-DI: 07045432088519; 2) LCSU 4, 300 ml, RTCA; Catalog Number: 880062; UDI-DI: 07045432088533; 3) LCSU 4, RTCA; Catalog Number: 881152; UDI-DI: 07045432067132;

View the original FDA recall notice

Why was this product recalled?

units manufactured in a limited time period may emit electromagnetic noise beyond the acceptable limits for RTCA application specified in RTCA DO-160G Chapter 21, Equipment Category M.

What actions did Laerdal Medical Corporation take?

On January 13, 2026, Medical Device Recall letters were sent to customers. Actions to be taken: 1. Please review this notice carefully and provide acknowledgment that you have received and understood the contents. If you received this notice from a distributor or other third party, please send your acknowledgment directly to Laerdal using the contact information provided below. You must provide the case number in the footer of this notice in your response. 2. Ensure that this notice is shared with all relevant personnel involved in the storage, maintenance, charging, and use of the LCSU 4 RTCA device. 3. Maintain this information for your records. As part of Laerdal s commitment to customer service, RTCA/DO-160G Section 21 Category M compliant replacement units are being offered, free of charge. To receive a replacement, you must reply within 2 weeks of receipt, with the following information: 1. Confirm acknowledgment of this notice as mentioned above. 2. Confirm serial numbers of your affected units, shipping address, and contact information in your reply. Once validated, Laerdal Medical will arrange for your affected unit(s) to be replaced at no cost. Important: Replacement units (881152) are shipped without a carry bag, battery, external power supply, wire stand, canister, or tubing. Please retain these accessories from your existing unit for continued use. Customers are required to confirm responsibility for disposing of the replaced LCSU 4 RTCA Unit(s) in accordance with applicable environmental regulations. Proof of disposal must be retained and made available upon request. By accepting the replacement, the customer releases Laerdal from any future claims or liabilities related to the old unit. Please maintain awareness of this notice and the related actions for your records. ***Update 03/24/2026***Additional recall letters were sent on or around 2/25/2026 to additional consignees identified by the recalling firm.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology