ATEC Lateral Navigation Disc Prep Instruments LIF NAV, OSTEOTOME 10mm REF 267-01-000-10-N UDI-DI code: 00190376541471; GMDN 12844 The ATEC Navigation Disc Prep Instruments are intended for use in surgical procedures. These instruments are non-sterile or sterile packaged, re-usable or single-use instruments designed to function with the Medtronic StealthStation System and NavLock Tracker. Recall
ATEC Lateral Navigation Disc Prep Instruments LIF NAV, OSTEOTOME 10mm REF 267-01-000-10-N UDI-DI code: 00190376541471; GMDN 12844 The ATEC Navigation Disc Prep Instruments are intended for use in surgical procedures. These instruments are non-sterile or sterile packaged, re-usable or single-use instruments designed to function with the Medtronic StealthStation System and NavLock Tracker.
FDA recall Class II
Ongoing
FDA lists this Class II recall for ATEC Lateral Navigation Disc Prep Instruments LIF NAV, OSTEOTOME 10mm REF 267-01-000-10-N UDI-DI code: 00190376541471; GMDN 12844 The ATEC Navigation Disc Prep Instruments are intended for use in surgical procedures. These instruments are non-sterile or sterile packaged, re-usable or single-use instruments designed to function with the Medtronic StealthStation System and NavLock Tracker.. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1343-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jan 2, 2026
- FDA report date
- Feb 18, 2026
- Quantity in commerce
- 15 units
- Distribution
- See key details
- FDA cause
- Device Design
What is being recalled?
ATEC Lateral Navigation Disc Prep Instruments LIF NAV, OSTEOTOME 10mm REF 267-01-000-10-N UDI-DI code: 00190376541471; GMDN 12844 The ATEC Navigation Disc Prep Instruments are intended for use in surgical procedures. These instruments are non-sterile or sterile packaged, re-usable or single-use instruments designed to function with the Medtronic StealthStation System and NavLock Tracker.
Affected model & UDI checker
Search the exact identifiers published for Z-1343-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 267-01-000-10-N | ATEC Lateral Navigation Disc Prep Instruments LIF NAV, OSTEOTOME 10mm REF 267-01-000-10-N UDI-DI code: 00190376541471; GMDN 12844 The ATEC Navigation Disc Prep Instruments are intended for use in surgical procedures. These instruments are non-sterile or sterile packaged, re-usable or single-use instruments designed to function with the Medtronic StealthStation System and NavLock Tracker. | Check source notice | Not specified |
| Not specified | ATEC Lateral Navigation Disc Prep Instruments LIF NAV, OSTEOTOME 10mm REF 267-01-000-10-N UDI-DI code: 00190376541471; GMDN 12844 The ATEC Navigation Disc Prep Instruments are intended for use in surgical procedures. These instruments are non-sterile or sterile packaged, re-usable or single-use instruments designed to function with the Medtronic StealthStation System and NavLock Tracker. | Check source notice | 00190376541471 |
FDA code information
Model/Catalog Number: 267-01-000-10-N UDI-DI code: 00190376541471 Lot Number: EM56693
Why was this product recalled?
Due a design issue where the navigated array connection geometry is incorrect.
What actions did Alphatec Spine, Inc. take?
On 01/02/2026, the firm emailed an "URGENT-MEDICAL DEVICE RECALL NOTIFICATION" Letter to customers informing them that Alphatec Spine has been notified of complaints regarding LIF Navigated Osteotome instruments appearing to have incorrect orientation when used with the Medtronic StealthStation". The instrument is correctly identified and displays the correct navigated length; however, the projected image on the Medtronic StealthStation" is rotated by 90. Customers are instructed to: "Review their inventory to determine if any of these affected devices are in their possession. If a device is within their possession, abstain from use and return the device to Alphatec Spine using the attached return label. "Share this notice within their organization or any organization where the affected device(s) has (have) been transferred. "Fill out the last page of this letter to confirm that they have read this notification and have taken all necessary actions, as described in this notification. For any questions/concerns or require assistance with the recall, e-mail [email protected] or call 1-800-922-1356.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1343-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology