Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1361-2026

Barco MNA with the HexaVue IP Integration System Recall

Barco MNA with the HexaVue IP Integration System, Model/Catalog Number: MNA-6x0-H

FDA recall Class II

Ongoing

FDA lists this Class II recall for Barco MNA with the HexaVue IP Integration System. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1361-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
May 23, 2025
FDA report date
Feb 18, 2026
Quantity in commerce
10 units
Distribution
See key details
FDA cause
Process control

What is being recalled?

Barco MNA with the HexaVue IP Integration System, Model/Catalog Number: MNA-6x0-H

Affected model & UDI checker

Search the exact identifiers published for Z-1361-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
MNA-6x0-HBarco MNA with the HexaVue IP Integration SystemCheck source noticeNot specified
Not specifiedBarco MNA with the HexaVue IP Integration SystemCheck source notice00724995220570
FDA code information

UDI-DI: 00724995220570; Serial Numbers for MNAs: 2531542441, 2531542470, 2531543549, 2531543611, 2531572419, 2531572433, 2531572440, 2531588558, 2531593768, and 2531595829

View the original FDA recall notice

Why was this product recalled?

This is a sub-recall of event RES 96885. The supplier identified that during their manufacturing process for a limited batch of MNA-6x0 (adapter) components the incorrect configuration file was used.

What actions did STERIS CORPORATION take?

All affected Customers have been contacted by a STERIS Service Technician arranging an onsite inspection of their unit(s) to locate and replace the affected MNA(s). Users are reminded to report any adverse events via FDA s MedWatch Database should an event occur. We apologize for any inconvenience this matter may cause, and as always, STERIS is dedicated to supporting our products and valued Customers. If you have questions regarding this matter, please contact Diksha Chechi, Quality Manager at 510-439-4500.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology