FDA RECALL RECORD · Z-1368-2026

Brand Name: Spectrum IQ Product Name: Infusion Pump Recall

Brand Name: Spectrum IQ Product Name: Infusion Pump Model/Catalog Number: 3570009 Software Version: N/A Product Description: Infusion Pump, one unit per carton Component: N/A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: Spectrum IQ Product Name: Infusion Pump. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1368-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 13, 2026
FDA report date
Feb 25, 2026
Quantity in commerce
5
Distribution
See key details
FDA cause
Employee error

What is being recalled?

Brand Name: Spectrum IQ Product Name: Infusion Pump Model/Catalog Number: 3570009 Software Version: N/A Product Description: Infusion Pump, one unit per carton Component: N/A

Affected model & UDI checker

Search the exact identifiers published for Z-1368-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
3570009Brand Name: Spectrum IQ Product Name: Infusion PumpCheck source noticeNot specified
Not specifiedBrand Name: Spectrum IQ Product Name: Infusion PumpCheck source notice00085412610900
FDA code information

Lot Code: GTIN 00085412610900, Serial Numbers 3815663, 3823468, 3824743, 3825234, and 3825599

View the original FDA recall notice

Why was this product recalled?

Pumps were released without full testing being performed, including occlusion alarm testing.

What actions did Baxter Healthcare Corporation take?

Firm notified the single customer on 1/13/26 via phone call. Customer was asked to work with on-site service technician to coordinate return of affected product.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology