Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1370-2026

Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit. Recall

Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1370-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Dec 31, 2025
FDA report date
Feb 25, 2026
Quantity in commerce
1,578 units
Distribution
See key details
FDA cause
Nonconforming Material/Component

What is being recalled?

Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.

Affected model & UDI checker

Search the exact identifiers published for Z-1370-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
7303Medica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.Check source noticeNot specified
Not specifiedMedica Capillary Tubes; EasyBloodGas/EasyStat Capillary Tube Kit.Check source notice00840095607575
FDA code information

Model No. 7303, UDI-DI: 00840095607575, Lot No. 25XXX.

View the original FDA recall notice

Why was this product recalled?

A systematic positive bias of up to 20% was identified across the reportable potassium range, resulting in false elevation results of patient samples when using Medica s capillary tubes.

What actions did Medica Corporation take?

An URGENT FIELD SAFETY NOTICE was sent to consignees on 01/09/25 informing them that potassium results obtained using affected devices in capillary mode may be falsely elevated. The FSN instructed consignees not to report potassium results obtained in capillary mode using affected lots on the Easy Stat Analyzer. The FSN is to be forwarded to end users and to where product was further distributed. Consignees are to contact their local representative with any questions.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.