Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1376-2025

Microstream Instructions for Use and Part Number used with - Product Description, REF: FilterLine H Set & VitaLine H Set Infant/Neonatal, 006147 - MICROSTREAM FilterLine H Set Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 006324; MICROSTREAM FilterLine H Set Long Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 007738; MICROSTREAM VitaLine H Set Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 010807. Recall

Microstream Instructions for Use and Part Number used with - Product Description, REF: FilterLine H Set & VitaLine H Set Infant/Neonatal, 006147 - MICROSTREAM FilterLine H Set Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 006324; MICROSTREAM FilterLine H Set Long Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 007738; MICROSTREAM VitaLine H Set Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 010807.

FDA recall Class II

Ongoing

FDA lists this Class II recall for Microstream Instructions for Use and Part Number used with - Product Description, REF: FilterLine H Set & VitaLine H Set Infant/Neonatal, 006147 - MICROSTREAM FilterLine H Set Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 006324; MICROSTREAM FilterLine H Set Long Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 007738; MICROSTREAM VitaLine H Set Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 010807.. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1376-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 17, 2025
FDA report date
Mar 26, 2025
Quantity in commerce
318292
Distribution
See key details
FDA cause
Labeling design

What is being recalled?

Microstream Instructions for Use and Part Number used with - Product Description, REF: FilterLine H Set & VitaLine H Set Infant/Neonatal, 006147 - MICROSTREAM FilterLine H Set Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 006324; MICROSTREAM FilterLine H Set Long Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 007738; MICROSTREAM VitaLine H Set Infant/Neonatal CO2 Sampling Line and Airway Adapter for Humid Environments, 010807.

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Part Number and Revision - REF/UDI-DI(Lot): 006147 Rev N - 006324/20884521551692(C210417257, C210957017, C211102266, C211206691, C220108497, D191171644, D191171646, D191176150, D191278690, D200316307, D200538149, D200639546, D200742524, D200745256, D200950569, D200951852, D201055941, D201261296, D210101027, D210103472, D210310524, D210413725, D210726722, D220211210), 10884521551695(D191171644, D191171646, D191176150, D191278690, D200316307, D200745256, D201261296, D210103472); 007738/20884521216010(D191069375, D191171650, D191176156, D200314292, D200431907, D200537959, D200742194, D200745606, D200950373, D201260700, D210308577, D210728936, D211047680, D211155459); 010807/20884521551920(D200101284, D200426597, D200742219, D200745037, D201260710, D210310546, D210837166), 10884521551923(D200101284, D200745037).

View the original FDA recall notice

Why was this product recalled?

Neonatal-Infant and Adult-Pediatric CO2 Sampling Line/Airway Adapter/CO2 Oral-Nasal Filter Line/Intubated Filter (CO2) Line device users may experience difficulty/inability to disconnect adapter from endotracheal tube for procedures (e.g. suctioning), may cause treatment delay and/or unintended extubation, which could lead to respiratory failure, hypoxia, aspiration, and/or low oxygen saturation.

What actions did Oridion Medical 1987 Ltd. take?

On 2/17/2025, correction notices were mailed to Risk Manager, Director of Respiratory Care, Director of Anesthesiology who were informed of the following: In September 2025, the firm will issue an update to the Instructions for Use (IFUs) for the The following IFU's: 01158, PT00156355, PT00156254, PT00156250 stating: Adverse events associated with attaching and detaching the airway adapter from the breathing circuit are listed in descending order of severity: unintended extubation, respiratory failure, hypoxia, low oxygen saturation, aspiration/inhalation and delay to treatment. Any serious incident related to device use that may occur should be reported immediately to the manufacturer, the local competent authority, and any other regulators as required. Pass on and post this notice for all those who need to be aware within your organization and to any organization where the affected product has been transferred or distributed. Complete and return the Customer Confirmation Form to [email protected] Adverse reactions or quality problems experienced with this product should be reported to the firm's Quality Assurance at 800-255-6774 option 1, then option 1. If you have questions contact the firm's Customer Service at 800-962-9888, Option 2. On 5/13/2025, additional correction notices identifying additional affected devices were mailed to additional customers.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology