Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula, REF: 7CN80R Recall
Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula, REF: 7CN80R
FDA recall Class I
Ongoing
FDA lists this Class I recall for Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula, REF: 7CN80R. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1390-2025
- Recall classification
- Class I
- Enforcement status
- Ongoing
- Initiation date
- Feb 26, 2025
- FDA report date
- Apr 2, 2025
- Quantity in commerce
- 780
- Distribution
- See key details
- FDA cause
- Under Investigation by firm
What is being recalled?
Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula, REF: 7CN80R
Affected model & UDI checker
Search the exact identifiers published for Z-1390-2025.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 7CN80R | Shiley Adult Flexible Tracheostomy Tube with TaperGuard Cuff Reusable Inner Cannula, REF: 7CN80R | Check source notice | Not specified |
FDA code information
UDI-DI/GTIN: A8845212054401, 20884521205441, 10884521205444. Lot: 202405258X. Expiration: 06/01/2029
Why was this product recalled?
Tracheostomy tube with cuff reusable inner cannula has a flange that may disconnect from the outer cannula of the device, which could cause respiratory failure, aspiration, unspecified tissue injury, aspiration, respiratory tract infection, bronchospasm, a delay to treatment and/or death.
What actions did Covidien take?
On 2/26/2025, recall notices were mailed and/or emailed to Risk Managers, Directors of Respiratory Care, Anesthesiology, Pulmonary/Intensive Care and ENT who were asked to do the following: 1) For patients with affected devices currently in place, a replacement is required. The patient's medical team should assess the overall patient risk when considering the timing of replacement. Clinicians should continue to follow current product Instructions For Use (IFU) along with facility specific policies and procedures. 2) Quarantine and return all unused product from the affected lot in your inventory to the firm. 3) Pass on this notice to all those who need to be aware within your organization or to any organization where the potentially affected product from the specified lot has been transferred or distributed. 4) Complete and return the enclosed Customer Confirmation Form via email to [email protected] If you have any questions regarding this communication, contact the firm's Customer Service at 800-962-9888, Option 2.
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1390-2025
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology