Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1464-2025

Medline procedure kits labeled as: 1) ACOUSTIC NEUROMA PACK-LF, Pack Number DYNJ37741B; 2) ANGIOGRAPHY PACK, Pack Number DYNJ21519W; 3) NEURO ANGIO PACK, Pack Number DYNJ83941; 4) NEURO ANGIO PACK SJF-LF, Pack Number DYNJ55678; 5) NEURO PACK, Pack Number DYNJ04851Q; 6) SNI NEURO, Pack Number DYNJ45379D; 7) VP SHUNT PACK, Pack Number DYNJ55558C Recall

Medline procedure kits labeled as: 1) ACOUSTIC NEUROMA PACK-LF, Pack Number DYNJ37741B; 2) ANGIOGRAPHY PACK, Pack Number DYNJ21519W; 3) NEURO ANGIO PACK, Pack Number DYNJ83941; 4) NEURO ANGIO PACK SJF-LF, Pack Number DYNJ55678; 5) NEURO PACK, Pack Number DYNJ04851Q; 6) SNI NEURO, Pack Number DYNJ45379D; 7) VP SHUNT PACK, Pack Number DYNJ55558C

FDA recall Class II

Ongoing

FDA lists this Class II recall for Medline procedure kits labeled as: 1) ACOUSTIC NEUROMA PACK-LF, Pack Number DYNJ37741B; 2) ANGIOGRAPHY PACK, Pack Number DYNJ21519W; 3) NEURO ANGIO PACK, Pack Number DYNJ83941; 4) NEURO ANGIO PACK SJF-LF, Pack Number DYNJ55678; 5) NEURO PACK, Pack Number DYNJ04851Q; 6) SNI NEURO, Pack Number DYNJ45379D; 7) VP SHUNT PACK, Pack Number DYNJ55558C. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1464-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 11, 2025
FDA report date
Apr 2, 2025
Quantity in commerce
422
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Medline procedure kits labeled as: 1) ACOUSTIC NEUROMA PACK-LF, Pack Number DYNJ37741B; 2) ANGIOGRAPHY PACK, Pack Number DYNJ21519W; 3) NEURO ANGIO PACK, Pack Number DYNJ83941; 4) NEURO ANGIO PACK SJF-LF, Pack Number DYNJ55678; 5) NEURO PACK, Pack Number DYNJ04851Q; 6) SNI NEURO, Pack Number DYNJ45379D; 7) VP SHUNT PACK, Pack Number DYNJ55558C

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

1) Pack Number DYNJ37741B: UDI/DI (each) 40193489682213 (case), Lot Number 20BDB113A; 2) Pack Number DYNJ21519W: UDI/DI (each) 40889942255611 (case), Lot Number 20BDB225A; 3) Pack Number DYNJ83941: UDI/DI (each) 40888277736574 (case), Lot Number 22JBA879Z; 4) Pack Number DYNJ55678: UDI/DI (each) 40193489468688 (case), Lot Number 22JBC551Z; 5) Pack Number DYNJ04851Q: UDI/DI (each) 40889942402176 (case), Lot Number 22JBC798Z; 6) Pack Number DYNJ45379D: UDI/DI (each) 40195327048748 (case), Lot Number 23FME560A; 7) Pack Number DYNJ55558C: UDI/DI (each) 40195327408214 (case), Lot Number 23LMG468A

View the original FDA recall notice

Why was this product recalled?

Medical convenience kits contain plastic syringes affected by the FDA Safety Alert issued on March 19, 2024. Leaks, breakage, and/other quality issues have been identified that may pose a risk to patient health.

What actions did MEDLINE INDUSTRIES, LP - Northfield take?

Medline Industries LP issued a recall notice to its consignees on 02/11/2025 via letter delivered via USPS and email. The notice explained the issue, provided a link to for information pertaining to the risk involved https://www.fda.gov/medical-devices/safety-communications/update-evaluating-plastic-syringes-made-china-potential-device-failures-fda-safety-communication) and request the syringes within the convenience be discarded, except for circumstances where no alternative syringes are readily available. If these syringes must be used, closely monitor for leaks, breakage, and any other problems, and report any quality issues to Medline and the FDA. Distributors were directed to notify their customers.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.