Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1465-2026

Convenience kits used for dialysis maintenance Medline ADD A CATH DIALYSIS KIT SKU ECVC8415A Centurion CENTRAL LINE INSERTION TRAY SKU DT19810 Centurion DIALYSIS BUNDLE WITH NO CATHETTER SKU CVI4310A Medline DIALYSIS CAP CHANGE KIT SKU DYNDC2425 Medline DIALYSIS CHANGE KIT SKU EBSI1453A Centurion HEMODIALYSIS ACCESS KIT SKU DT12970 Medline LARGE BORE, DIALYSIS, ENHANCED SECUREMENT SKU EBSI1741 Recall

Convenience kits used for dialysis maintenance Medline ADD A CATH DIALYSIS KIT SKU ECVC8415A Centurion CENTRAL LINE INSERTION TRAY SKU DT19810 Centurion DIALYSIS BUNDLE WITH NO CATHETTER SKU CVI4310A Medline DIALYSIS CAP CHANGE KIT SKU DYNDC2425 Medline DIALYSIS CHANGE KIT SKU EBSI1453A Centurion HEMODIALYSIS ACCESS KIT SKU DT12970 Medline LARGE BORE, DIALYSIS, ENHANCED SECUREMENT SKU EBSI1741

FDA recall Class II

Ongoing

FDA lists this Class II recall for Convenience kits used for dialysis maintenance Medline ADD A CATH DIALYSIS KIT SKU ECVC8415A Centurion CENTRAL LINE INSERTION TRAY SKU DT19810 Centurion DIALYSIS BUNDLE WITH NO CATHETTER SKU CVI4310A Medline DIALYSIS CAP CHANGE KIT SKU DYNDC2425 Medline DIALYSIS CHANGE KIT SKU EBSI1453A Centurion HEMODIALYSIS ACCESS KIT SKU DT12970 Medline LARGE BORE, DIALYSIS, ENHANCED SECUREMENT SKU EBSI1741. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1465-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 8, 2026
FDA report date
Mar 4, 2026
Quantity in commerce
14,525
Distribution
See key details
FDA cause
Process control

What is being recalled?

Convenience kits used for dialysis maintenance Medline ADD A CATH DIALYSIS KIT SKU ECVC8415A Centurion CENTRAL LINE INSERTION TRAY SKU DT19810 Centurion DIALYSIS BUNDLE WITH NO CATHETTER SKU CVI4310A Medline DIALYSIS CAP CHANGE KIT SKU DYNDC2425 Medline DIALYSIS CHANGE KIT SKU EBSI1453A Centurion HEMODIALYSIS ACCESS KIT SKU DT12970 Medline LARGE BORE, DIALYSIS, ENHANCED SECUREMENT SKU EBSI1741

The retrieved code information does not contain explicit model/GTIN pairs that this publication can reliably map.

Search FDA identifier text

Search a model, catalog, lot or UDI value exactly as printed in FDA’s product description or code information.

This search checks only the FDA product description and code information displayed on this page. It does not decode a UDI, normalize GTIN lengths or establish a relationship between separately listed codes.

FDA code information

Medline ADD A CATH DIALYSIS KIT SKU ECVC8415A UDI-DI (ea) 10653160383628 UDI-DI (case) 00653160383621 Kit lots 2024121390 2025033190 Centurion CENTRAL LINE INSERTION TRAY SKU DT19810 UDI-DI (ea) 10653160368755 UDI-DI (case) 00653160368758 Kit lot 2025041090 Centurion DIALYSIS BUNDLE WITH NO CATHETTER SKU CVI4310A UDI-DI (ea) 10653160374978 UDI-DI (case) 00653160374971 Kit lots 2024051490 2025021290 2025051390 2025091090 Medline DIALYSIS CAP CHANGE KIT SKU DYNDC2425 UDI-DI (ea) 10653160375845 UDI-DI (case) 00653160375848 Kit lots 2024060690 2024101690 2024103190 2025010890 2025040190 Medline DIALYSIS CHANGE KIT SKU EBSI1453A UDI-DI (ea) 10653160383475 UDI-DI (case) 00653160383478 Kit lots 2024050790 2024051390 2024082290 2024111990 2025013090 2025032090 2025050190 2025062790 Centurion HEMODIALYSIS ACCESS KIT SKU DT12970 UDI-DI (ea) 10653160368922 UDI-DI (case) 00653160368925 Kit lots 2025040401 2025081401 Medline LARGE BORE, DIALYSIS, ENHANCED SECUREMENT SKU EBSI1741 UDI-DI (ea) 10653160374527 UDI-DI (case) 00653160374520 Kit lots 2024061780 2024070180 2024081280 2024111290 2025012490 2025032690 2025051390 2025061790 2025071490 2025072190

View the original FDA recall notice

Why was this product recalled?

Affected lots may exhibit issues with the silicone seal on Tego Connectors included in kits. Issues include silicone seal doming and silicone seal tearing. These issues may result in an occluded fluid path identified by the inability of the user to inject or withdraw blood product with a syringe. Damaged seals may also result in delay in therapy or fluid leakage, interruption in therapy, exposure to biological contaminants or air infused into the body.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Medline Industries, LP take?

On January 8, 2026, the firm notified affected customers of the product issue through email and first class mail. Customers were instructed to request stickers to over-label affected convenience kits with a warning label indicating to remove the affected component and replace it with product from supply. No kits are to returned.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.