Adapt Pump Recall
Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure
FDA recall Class II
Ongoing
FDA lists this Class II recall for Adapt Pump. Review the affected identifiers, reported reason and manufacturer instructions below.
Recall at a glance
- Recall number
- Z-1472-2026
- Recall classification
- Class II
- Enforcement status
- Ongoing
- Initiation date
- Jan 21, 2026
- FDA report date
- Mar 4, 2026
- Quantity in commerce
- 4286
- Distribution
- See key details
- FDA cause
- Software design
What is being recalled?
Adapt Pump, Model/Catalog Number: 61600200-Adapt Pump, component of Adapt support surfaces (Adapt Air, Adapt Air EX, Adapt Air Pro, Adapt Air Pro EX, Adapt Convertible, Adapt Convertible EX), sofware version 1.3.0; Mattress, Air Flotation, Alternating Pressure
Affected model & UDI checker
Search the exact identifiers published for Z-1472-2026.
The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.
Affected products
| Model / catalog number | Product | Lot scope reported | UDI / GTIN |
|---|---|---|---|
| 61600200-Adapt | Adapt Pump | Check source notice | Not specified |
| Not specified | Adapt Pump | Check source notice | 0084569904914 |
FDA code information
UDI-DI 0084569904914; All serial numbers with a manufacturing date prior to 12/04/2025
Why was this product recalled?
Agiliti identified an issue where the Adapt Pump (61600200-Adapt Pump) supplied with all Adapt Line support surfaces does not re-engage the Microclimate Management function after using the Autofirm function for 10 minutes. This issue only occurs when the 10-minute timer is allowed to elapse after the Autofirm button is pressed after initial setup, and it only affects the Microclimate Management function.
What actions did Agiliti Health - Ellis take?
The firm issued an URGENT: Medical Device Correction notice to its consignees on 01/21/2026 via email. The notice explained the issue with the device, potential risk, how to recognize the failure, and requested the following: "Actions to be Taken by Customer 1. Immediately share this Notice with all necessary parties within your organization. 2. Examine the stock in your organization to determine if you have any of the devices in your organization. a. If any devices have been transferred to another organization, please forward a copy of this Advisory Notice and the attached Customer Response Form to that organization. 3. Complete the attached Customer Response Form, even if your organization no longer possesses any devices. 4. Follow the instructions below when using Adapt Line support surfaces to ensure the Microclimate Management function is not inadvertently turned off. Temporary actions Until Agiliti is able to install updated software on devices in your possession to correct the issue, please do the following to ensure that the Microclimate function is restored after using Autofirm. After using Autofirm (if the 10-minute timer elapses) 1. Press the Microclimate Management button to toggle the function on. 2. Verify the indicator light next to the Microclimate Button is illuminated. 3. Verify all previous therapy settings have been restored." For questions regarding this Notice or the Recalled Device, please contact your local Service Representative or the Agiliti customer care team: " Email: [email protected] " Phone: 800-814-9389, option 3
Read the manufacturer action recorded by FDARecall timeline
Recall initiated by firm
FDA notice posted
FDA classification recorded
FDA enforcement report date
Related FDA submissions
No submission number with an exact verified relationship is linked in the retrieved recall notice.
FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.
Source and update history
- FDA recall notice Z-1472-2026
- openFDA enforcement record · Source release Nov 11, 2026 · Retrieved Nov 21, 2026
- openFDA device recall record · Source release Nov 19, 2026 · Retrieved Nov 21, 2026
- FDA enforcement data
First stored snapshot of this source.
- FDA device-recall notice
First stored snapshot of this source.
The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.
Our data sources and methodology