FDA lists this Class II recall for Nasopore Ex Firm 8cm fragmentable nasal dressing, Catalog Number 5400-030-008 and Nasopore Forte plus 8cm Catalog Number 5400-030-008ITL. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1479-2025
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Feb 25, 2025
FDA report date
Apr 9, 2025
Quantity in commerce
182,344 total units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Nasopore Ex Firm 8cm fragmentable nasal dressing, Catalog Number 5400-030-008 and Nasopore Forte plus 8cm Catalog Number 5400-030-008ITL

Affected model & UDI checker

Search the exact identifiers published for Z-1479-2025.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
5400-030-008Nasopore Ex Firm 8cm fragmentable nasal dressing, Catalog Number 5400-030-008 and Nasopore Forte plus 8cm Catalog Number 5400-030-008ITLCheck source noticeNot specified
5400-030-008ITLNasopore Ex Firm 8cm fragmentable nasal dressing, Catalog Number 5400-030-008 and Nasopore Forte plus 8cm Catalog Number 5400-030-008ITLCheck source noticeNot specified
Not specifiedNasopore Ex Firm 8cm fragmentable nasal dressing, Catalog Number 5400-030-008 and Nasopore Forte plus 8cm Catalog Number 5400-030-008ITLCheck source notice07613154340842
Not specifiedNasopore Ex Firm 8cm fragmentable nasal dressing, Catalog Number 5400-030-008 and Nasopore Forte plus 8cm Catalog Number 5400-030-008ITLCheck source notice07613327359473
FDA code information

5400-030-008 UDI-DI: 07613154340842, Lot Numbers: 2023012515 2023012715 2023022012 2023050814 2023072632 2023102312 2023102612 2024032514 2024040212; 5400-030-008ITL UDI-DI: 07613327359473, Lot Numbers: 2022092714 2023020633 2023032433 2023050915 2023052514 2023062216 2023082814 2023101112 2024031812 2022092914 2023020814 2023032814 2023051114 2023053115 2023062032 2023090618 2023101714 2024042212 2022101332 2023020915 2023033016 2023051215 2023060914 2023082215 2023100312 2023102014 2024042312 2022121932 2023021416 2023050115 2023052214 2023061412 2023082312 2023100412 2024030414 2024061814 2022122032 2023022032 2023050515 2023052412 2023061514 2023082514 2023100912 2024030512 2024062814

View the original FDA recall notice

Why was this product recalled?

There is a potential for blister seals on the product to present a bubble on the seal area of the packaging. This seal acts as a sterile barrier and, in this event, indicates that the sterility seal has been breached.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did Stryker Corporation take?

An Urgent Medical Device Recall notification letter dated 2/24/25 was sent to customers. Actions to be taken 1. Immediately review your inventory to locate and quarantine any affected products at your facility. 2. Return the enclosed Business Reply Form (BRF), even if the affected product is no longer in inventory, to the email [email protected]. 3. Upon receipt of the completed BRF, Stryker will provide a shipping label to return recalled product on-hand. 4. A replacement will be provided upon receipt of the recalled product. 5. Maintain awareness of this communication internally and inform Stryker if any of the subject devices have been distributed to other organizations. If so, provide contact details so Stryker can inform the recipients accordingly. [email protected]. RESPONSE IS REQUIRED.

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology