Medical devices, healthcare technology and the business of care
FDA RECALL RECORD · Z-1485-2026

Brand Name: AESCULAP Product Name: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Recall

Brand Name: AESCULAP Product Name: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Model/Catalog Number: FF399R Software Version: N/A Product Description: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Component: N/A

FDA recall Class II

Ongoing

FDA lists this Class II recall for Brand Name: AESCULAP Product Name: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM. Review the affected identifiers, reported reason and manufacturer instructions below.

Recall at a glance

Recall number
Z-1485-2026
Recall classification
Class II
Enforcement status
Ongoing
Initiation date
Jan 15, 2026
FDA report date
Mar 11, 2026
Quantity in commerce
126 units
Distribution
See key details
FDA cause
Under Investigation by firm

What is being recalled?

Brand Name: AESCULAP Product Name: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Model/Catalog Number: FF399R Software Version: N/A Product Description: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MM Component: N/A

Affected model & UDI checker

Search the exact identifiers published for Z-1485-2026.

The lookup covers structured identifiers listed on this page. A full UDI may also contain lot or serial information; a GTIN match alone does not check those conditions.

Affected products

Model / catalog numberProductLot scope reportedUDI / GTIN
FF399RBrand Name: AESCULAP Product Name: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MMCheck source noticeNot specified
Not specifiedBrand Name: AESCULAP Product Name: MINOP TROCAR 150MM 4 WKING CHANNEL6.0MMCheck source notice04038653065547
FDA code information

Mode No. FF399R; UDI-DI 04038653065547; Serial Numbers: 4595, 4619, 4657, 4764, 4592, 4596, 4598, 4601, 4606, 4607, 4608, 4612, 4616, 4618, 4620, 4621, 4623, 4659, 4661, 4662, 4667, 4670, 4673, 4674, 4688, 4689, 4746, 4748, 4757, 4758, 4760, 4761, 4762, 4767, 4768, 4771, 4773, 4774, 4779, 4780, 4784, 4786, 4789, 4791, 4792, 4795, 4799, 4811, 4812, 4818, 4819, 4821, 4822, 4824, 4825, 4826, 4827, 4836, 4838, 4841, 4842, 4843, 4854, 4856, 4858, 4864, 4868, 4869, 4880, 4882, 4886, 4890, 4891, 4892, 4893, 4899, 4901, 4902, 4907, 4914, 4921, 4927, 4940, 4951, 4959, 4967, 4968, 4970, 4971, 4972, 4973, 4974, 4975, 4976, 4990, 4994, 4995, 5000, 5004, 5015, 5024, 5035, 5036, 5042, 5057, 5058, 5059, 5062, 5063, 5064, 5072, 5076, 5081, 5084, 5106, 5112, 5115, 5117, 5118, 5119, 5121, 5123, 5165, 5166, 5168, 5172;

View the original FDA recall notice

Why was this product recalled?

There is the potential for the length of the trocar shaft to be too long.

These are potential consequences described in the source. The recall reason does not establish how many injuries or incidents occurred.

What actions did AESCULAP, INC. take?

on January 28, 2026 URGENT MEDICAL DEVICE CORRECTION NOTIFICATION letters were sent to customers. Actions to be taken: - Review the attachment Aesculap AG Urgent Field Safety Notice in its entirety. - If it is determined that a portion of the visual field is obscured, the manufacturer recommends rotating the device to compensate for the restricted view (refer to the attached Urgent Field Safety Notice from the manufacturer). Note: Devices are not being removed from the market, this is a field notification. - Return the completed Urgent Medical Device Correction Acknowledgement Form to Aesculap, Inc. Quality Assurance department by faxing the form to (610) 849-1197 or e-mail to [email protected] within two (2) weeks of receipt. Action planned by Aesculap AG to correct the problem: To address this issue, the manufacturer is implementing additional inspections of the optical system and introducing clearly defined acceptance criteria, along with other corrective actions. Should there be any questions, comments, or concerns regarding the product you have received, please feel free to contact (844-903-6417) or [email protected]

Read the manufacturer action recorded by FDA

Recall timeline

  1. Recall initiated by firm

  2. FDA notice posted

  3. FDA classification recorded

  4. FDA enforcement report date

Related FDA submissions

No submission number with an exact verified relationship is linked in the retrieved recall notice.

FDA’s notice explicitly links the submission numbers shown here. Read each filing for its own recorded device and scope.

Source and update history

  1. FDA enforcement data

    First stored snapshot of this source.

  2. FDA device-recall notice

    First stored snapshot of this source.

The history starts with the versions held by MedtechLedger. Earlier source changes are not reconstructed. Official dates and recorded status come from FDA; page summaries describe the source information.

Our data sources and methodology